Franconia FC 02 Скачать руководство пользователя страница 10

6.0 Software operation

Overview

After installation of the software and configuration of the hardware devices you are ready to

Double-click on 

FC02.exe

 (Directory 

FC02/FC02.exe

)

The following main menu is displayed:

The FC 02 software checks automatically the status of the network connected on the optic fiber link, and 
shows 2 main parts in its standard version.
The upper part concerns the FTM (Frankonia Turntable), and the lower part concerns the FAM 
(Frankonia antenna mast).
When a device is not found, the software indicates it, and the corresponding part is shown in grey.
When you power on the system, it needs to be initialized in order to find its reference point.
The initialisation procedure is performed automatically and described in the FTM and FAM user manuals.
The software indicates automatically if turntable and/or antenna mast have to be initialized.
How to start the initialisation run for turntable and antenna mast is described later in this manual.

MA_FC02_(Rev2_1) - 9/06 / Version 1 / Rev:2.1

10 / 24

Содержание FC 02

Страница 1: ...USER MANUAL FC 02 Controller for turntable and antenna mast MA_FC02_ Rev2_1 9 06 Version 1 Rev 2 1 1 24...

Страница 2: ...7 5 0 Software installation 9 Hardware and Software Requirements 9 Installation of FC 02 application software 9 6 0 Software operation 10 Overview 10 Turntable controls 11 Antenna mast controls 13 Pr...

Страница 3: ...he available line voltage and that the correct fuse is installed Ground the instrument This product is a Safety Class 1 Product provided with a protective earthing ground incorporated in the power cor...

Страница 4: ...t ways of control by providing serial RS 232 as well as GPIB IEEE488 2 interfaces The connection to the different devices is made using optic fibers duplex cables in order to avoid any RF interference...

Страница 5: ...l as a detailed description of this procedure This offers also the possibility to modify easily our positioning controller in order to solve some communication problems with some testing software Fron...

Страница 6: ...ruments SCPI and the IEEE 488 2 protocol on which SCPI is based 5 Serial connector RS 232 The RS 232 port is a standard serial port Baud rate 9600 used for the manual mode or used for upgrading the fl...

Страница 7: ...lex fiber starts from the FC 02 Connections to computer Several possibilities are offered for connecting the FC 02 to the Computer in charge of the instrumentation control The FC 02 can be used in man...

Страница 8: ...everal windows could be open on the computer PC1 Also orders can be given manually to the FC 02 using our software Automatic mode with instrumentation The FC 02 could be connected also to 2 separate c...

Страница 9: ...d Software Requirements You should have the following hardware and software to work with your FC 02 PC with Windows NT 2000 XP operating system compatible mouse and VGA graphics min 800 x 600 pts Inst...

Страница 10: ...Turntable and the lower part concerns the FAM Frankonia antenna mast When a device is not found the software indicates it and the corresponding part is shown in grey When you power on the system it n...

Страница 11: ...movement of antenna mast in m Offset Enter offset of movement of antenna mast in m Start init Start an initialisation run of the antenna mast Program Load settings Strg L Load settings Save settings S...

Страница 12: ...rom 200 to 200 depending on the settings for offset limits Step from 0 to 200 Speed from 1 slow to 30 fast Please notice The rotating knob represents the physical position of the turntable and is not...

Страница 13: ...the antenna mast can be changed if needed As a factory setting these values are set to 0 8m and 4m in case of 4m antenna mast or 6m in case of 6m antenna mast We have defined these values in order to...

Страница 14: ...not influenced by the settings of limits and offset The displayed values m take into account the setting of offset If the software detects a motion initiated by an external software all buttons excep...

Страница 15: ...Program functions Each time you start the software the file default set is loaded The default settings of this file are explained in the table below MA_FC02_ Rev2_1 9 06 Version 1 Rev 2 1 15 24...

Страница 16: ..._h2 0 000 m Antenna mast 2 horizontal offset offset_v2 0 000 m Antenna mast 2 vertical offset limit_h_max2 4 000 m Antenna mast 2 hor limit limit_h_min2 0 800 m Antenna mast 2 hor limit limit_v_max2 4...

Страница 17: ...after completed execution of all previous commands OPC Query Operation Complete reports an additional 1 after completed execution of all previous commands RST Sets the equipment to initial state Limit...

Страница 18: ...Setting of the operation enable register STAT QUES EVEN Query only Output of the questionable register STAT QUES COND Query only Output of the questionable condition register STAT QUES ENAB 0 65535 Se...

Страница 19: ...AX deg POS X ANGL LIM LOW NRf 200 0 MIN MAX deg Upper and lower limit value of the position POS X ANGL OFFS NRf 200 MIN MAX deg Offset of the position The limit values remain unchanged POS X ANGL VEL...

Страница 20: ...NRf value Numerical entries are admissible optionally with prefix and exponent Value is the valid numerical range The value can be followed by a unit For indication of lengths mm cm and m are valid f...

Страница 21: ...ependent Error equipment error 4 Execution Error execution error 5 Command Error command error 6 not used always 0 7 Power on Controller was switched on Operation Status Register OPER 0 Calibration su...

Страница 22: ...150 String data error Executing error Error code Error Description 200 Execuiting error 213 Init ignored 221 Setting conflict 222 Data out of range 224 Illegal parameter value 240 Hardware error Devic...

Страница 23: ...Programming Languages SCPI Remote Interface GPIB IEEE488 2 and RS 232 Baud rate 9600 Optical Input Output 660 nm FSMA connectors Accessories included Power cord application software User Manual Notic...

Страница 24: ...MA_FC02_ Rev2_1 9 06 Version 1 Rev 2 1 24 24 FRANKONIA GmbH Industriestra e 16 91180 Heideck Germany TEL 49 0 9177 98500 FAX 49 0 9177 98520 info Frankonia EMC com www Frankonia EMC com...

Отзывы: