8. Troubleshooting
8.4.3 Troubleshooting CAN Sub-Object 0x2001Sub1
Table 48: Troubleshooting CAN Sub-Object 0x2001Sub1
Problem
Possible cause
Solution
Bit 1 of
0x2001sub1
set.
See
.
Initializing
u
Wait until the initialization is
complete.
u
Reboot the Blackbird OECD. Turn the
power supply off and on.
Bit 2 of
0x2001sub1
set.
See
.
Rebooting soon
u
Wait until the reboot is complete.
Bit 3 of
0x2001sub1
set.
See
.
Self Diagnosis
u
Wait until the automatic
troubleshooting is complete.
u
Reboot the Blackbird OECD. Turn the
power supply off and on.
Bit 4 of
0x2001sub1
set.
See
.
Webserver Error
u
Reboot the Blackbird OECD. Turn the
power supply off and on.
Bit 7 of
0x2001sub1
set.
See
.
Light Source
Error
u
Reboot the Blackbird OECD. Turn the
power supply off and on.
Bit 8 or 11 of
0x2001sub1
set.
See
.
Temperature
Warning
Temperature
Error
u
Reboot the Blackbird OECD. Turn the
power supply off and on.
u
Make sure to operate within
specification. See
.