background image

8. Troubleshooting

Signal Quality

Status codes

Solution

Wavelength Too Low

or

Wavelength Too

High

u

Make sure that connected
sensors are within the
measuring range. See

Measurement and Sensors,
page 12

.

u

Make sure that only permitted
sensors are connected. See

Measurement and Sensors,
page 12

.

Device Error

u

Reboot the Blackbird OECD.
Turn the power supply off and
on.

Calibration

Missing

u

Contact fos4X Support. See

fos4X Support, page 73

.

No Signal

u

Make sure that a sensor is
connected.

u

Clean the optical connectors.
See 

Cleaning E-2000/APC

Connections, page 68

.

u

Reboot the Blackbird OECD.
Turn the power supply off and
on.

Device

Initializing

u

Wait until the initialization is
complete.

8.3 Troubleshooting Ethernet

TIP

If the described options do not help to fix the problem:

u

Contact fos4X Support. See 

fos4X Support, page 73

.

62 / 80

MDE0835_4-10_1.3.x_5.0_Blackbird

OECD_(6CH, CNO)-en-us

© fos4X 2020

Содержание Blackbird OECD MDE0835

Страница 1: ...ype Measuring device Product Version 1 3 x Document Version 5 0 Issue Date 2020 02 07 Language en us INSTRUCTIONS FOR USE Blackbird OECD 6CH CNO MDE0835 Product Type Measuring device Product Version 1 3 x Document Version 5 0 Issue Date 2020 02 07 Language en us ...

Страница 2: ...d Violations will result in the liability for damages All rights reserved in the event of patent utility model or design registration All rights reserved fos4X is a trademark of fos4X GmbH Brand names and protected trademarks are the property of their respective owners The naming of brand names and protected trademarks has only descriptive character Named trademarks are not part of any partnership...

Страница 3: ...ation of Personnel 9 2 4 Intended Use 10 2 5 Residual Risks 10 2 5 1 Explosion Protection 10 2 5 2 Working Environment Wind Turbine 10 3 Product Description 11 3 1 Functional Description 11 3 1 1 Overview 11 3 1 2 Operation 12 3 2 Technical Data 12 3 2 1 Measurement and Sensors 12 3 2 2 Interfaces 14 3 2 3 Optical Data 14 3 2 4 Electrical Data 15 3 2 5 Environmental Conditions 15 3 2 6 Tested Stan...

Страница 4: ...4 2 Packaging 47 4 3 Unpacking 47 4 4 Packing 48 4 5 Transport 48 4 6 Storage 49 5 Installation 50 5 1 Important Safety Notes for Installation 50 5 2 Attaching the Measuring Device in the Control Cabinet 50 5 3 Connecting the Interfaces 50 5 4 Connecting the Power Supply 51 5 5 Connecting a Sensor 52 6 Commissioning 54 6 1 Commissioning the Ethernet Interface 54 6 2 Commissioning the CAN Interface...

Страница 5: ...bleshooting CAN Sub Object 0x2001Sub1 66 9 Maintenance 68 9 1 Preventive Maintenance 68 9 2 Corrective Maintenance 68 9 2 1 Cleaning E 2000 APC Connections 68 9 2 2 Replacing the PWR Connector 69 9 2 3 Replacing the CAN Connector 70 9 2 4 Firmware Update 70 9 3 fos4X Support 73 9 4 Spare Parts and Accessories 74 10 Taking out of Service and Disposal 75 10 1 Taking out of Service 75 10 2 Disassembl...

Страница 6: ...Safety chapter applies to the entire use of the product See Safety page 9 Safety notes at the beginning of a section apply to the entire section Safety notes before a step apply to the entire following section Safety notes are shown in the operating instructions as follows DANGER Nature and cause of risk Consequences u Measures to avoid the risk The safety notes with the signal words Caution and N...

Страница 7: ... of an instruction List Point 1 Point 2 Enumeration with title Identification of Information page 6 Cross reference Bold Text Highlighting of particularly important information Software text Text parameter or input in the software WINDOW Name of a window or button in the software CONTROL ELEMENT Designation of a control element of the product 1 4 Abbreviations CAN Controller Area Network CNO CANop...

Страница 8: ...D Light Emmiting Diode Max Maximum MTBF Mean Time Between Failures NMT CANopen Network Management OECD Opto Electronic Converter Device PC Personal Computer SMF Single Mode Fiber SNTP Simple Network Time Protocol UDP User Datagram Protocol 8 80 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us fos4X 2020 ...

Страница 9: ...Trained personnel Personnel trained in the operation and safe use of the product Qualified personnel Personnel who based on their specialist training knowledge and experience and knowledge of the relevant regulations are able to assess the work assigned to them and identify possible hazards Assignment of tasks Tasks within the scope of use may only be carried out by the personnel defined below Tab...

Страница 10: ... Residual Risks 2 5 1 Explosion Protection The product is not suitable for use in potentially explosive environments 2 5 2 Working Environment Wind Turbine The use of the product in wind turbines requires special risk prevention measures Measures for risk prevention on wind turbines u Obtain and observe safety and work regulations Qualification and training of personnel Personal protective equipme...

Страница 11: ...lackbird OECD consists of the following components Mounting plate Used to attach the Blackbird OECD with 4 M6 bolts or welding studs E 2000 APC connectors 6 sockets for the connection of fiber optical sensors with E 2000 APC connector Power supply Figure 1 PWR connector PWR MC 1 5 2 GF 3 81 connector for the power supply with 10 28 V DC Interfaces Figure 2 ETH connector ETH RJ45 connector for the ...

Страница 12: ...gnals of the sensors and provides the wavelengths of the individual sensors via the following interfaces Ethernet CANopen 3 2 Technical Data 3 2 1 Measurement and Sensors Measuring Properties Table 3 Measuring Properties Property Value Unit Channels 6 Wavelength stability over the temperature range 20 pm Noise1 1 pm Resolution2 0 01 pm Warm up time3 20 min 3 dB cut off frequency 500 Hz Internal sa...

Страница 13: ...tion 4 Triggered by CANopen SYNC messages Requirements for sensors Table 4 Requirements for sensors Property Value Unit Sensor type Fiber Bragg grating Reflectivity 70 20 Full width half maximum FWHM 400 600 pm Side lobe suppression 12 dB Compatible sensors from fos4X Table 5 Compatible sensors from fos4X Measurand Sensor Strain fos4Strain fos4Strain expert Vibration fos4Acc 1D fos4Acc 2D fos4Acc ...

Страница 14: ...N Table 7 CAN Interface Property Value Unit Connector MC 1 5 3 GF 3 81 Max CAN data rate 1 Mbps Output rate 100 Hz Protocol CANopen 3 2 3 Optical Data Table 8 Optical Data Property Value Unit Connector type E 2000 APC Fiber type SMF 28 compatible Output power per channel max 10 mW Output power per channel typical 0 5 1 mW Laser Class DIN EN 60825 1 1 14 80 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6C...

Страница 15: ...tection class IP20 Maximum operating altitude 3000 m Permitted temperature Transport and storage 40 85 C Permitted relative humidity Non condensing Transport and storage 5 95 Permitted temperature Operation 20 65 C Permitted relative humidity Non condensing Operation 5 95 Cooling Passive The specified protection class only applies if the cable connections have the same or a higher protection class...

Страница 16: ... 3M6 IEC 60068 2 2 Class 3M6 Electromagnetic Compatibility IEC 61326 1 EN 55032 MTBF Table 12 MTBF Property Value Unit MTBF 100000 h 3 2 7 Mechanical Data Table 13 Mechanical Data Property Value Unit Length 186 mm Height 141 mm Width 148 mm Total weight 1 6 kg Blackbird OECD without packaging Table 14 Mounting Property Value Unit Mounting 4x M6 16 80 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO ...

Страница 17: ...code of the serial number CE marking RoHS marking Indication of the laser class of the product Class 1 Laser Product Power supply 12 24V Power consumption max 15W Optional Device designation KF123 Optional Customer specific ERP number ERP 123456 GTIN P N 4251112345678 3 4 Display and Control Elements 3 4 1 Status LEDs 3 4 1 1 LED Flashing Patterns The Blackbird OECD uses different LED flashing pat...

Страница 18: ...tion On LED is continuously on Slow flashing 1 s Fast flashing 1 s Single flash 1 s Double flash 1 s Triple flash 1 s Quadruple flash 1 s 3 4 1 2 LEDs on the PWR connector There are 2 LEDs on the PWR connector 18 80 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us fos4X 2020 ...

Страница 19: ...on the PWR connector STATUS LED Green Yellow Red On System healthy Measurement warning Measurement error Slow flashing Sytem auto adjust System error Fast flashing Identify System not adjusted TIP The STATUS LED flashes quickly every 4 s when in the state On Table 18 STATUS LED and device state Device state Description System healthy Normal operation fos4X 2020 MDE0835_4 10_1 3 x_5 0_Blackbird OEC...

Страница 20: ...r Identify Status is activated by a network command The blink pattern can be used to identify the Blackbird OECD System error Error in the power supply Error in the light source General system error System not adjusted No adjustment for the OE chips loaded Microcontroller defective 3 4 1 3 LEDs on the ETH Connector There are 2 LEDs on the Ethernet connector Figure 6 LED on the Ethernet connector 1...

Страница 21: ...N Connector 1 Switch for terminating resistor 2 Run LED green 3 Error LED red 4 LED not in use Switch for terminating resistor The CAN connector has a 120 Ohm terminating resistor The terminating resistor can be activated and deactivated via the switch 1 In the right position the terminating resistor is active Run LED Table 20 Run LED on the CAN Connector State Device state Description On Operatio...

Страница 22: ... Warning limit reached At least one error counter of the CAN controller has reached or exceeded the warning level Double flash Error control event A guard event NMT slave or NMT master or a heartbeat event heartbeat consumer has occured Triple flash Sync error The sync message was not received within the configured time span for the communication cycle On Bus off Error with the connection to the C...

Страница 23: ...The web interface can be accessed in a web browser within the network via the IP address of the Blackbird OECD Factory setting 192 168 39 99 The web interface is available via the following interfaces Ethernet The web interface has the following functions Monitoring of the system status of the Blackbird OECD and the connected sensors Updating the firmware of the Blackbird OECD Setting the configur...

Страница 24: ...t the system state CAN Status of the CAN bus Signals Informations about the measurement channels Uptime Operating time of the system System Information Informations about the system Device Type Type designation and version of the Blackbird OECD Serial Number Serial number of the Blackbird OECD Firmware Version Currently installed firmware version 24 80 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CN...

Страница 25: ... 5 2 2 System State The page SYSTEM STATE consists of the following pages CAN STATE SIGNAL STATE HARDWARE STATE SYSTEM REBOOT CAN State Figure 9 CAN State fos4X 2020 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us 25 80 ...

Страница 26: ...AN Module State Status of the CAN module CAN Diagnostic Counters Counters of the CAN bus for diagnose purposes Signal State Figure 10 Signal State number of channels depends on the variant Table 25 Signal State Informations about each channel Column Content Channel Number of the channel Wavelength Wavelength of the sensor Raw Signal Signal value of the sensor Reference top and Filter bottom State ...

Страница 27: ...ate Informations about the CAN interface Column Content Node State Status of the CANopen node CAN Bus State Status of the CAN bus Serial number Serial number of the Blackbird OECD Software State Status of the software Software Software Version Temp Temperature fos4X 2020 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us 27 80 ...

Страница 28: ...re Software Version Temp Temperature LS Number of the light source connected to the channel Table 28 Light Source States Informations about the light source Column Content Number of the light source Laser Power Current power of the laser in percent Laser State Status of the laser of the light source Currents Current of the laser top and of the temperature regulation of the laser bottom Software St...

Страница 29: ...tion System Reboot Figure 12 System Reboot A restart of the Blackbird OECD system can be carried out by clicking on the PERFORM SYSTEM REBOOT button fos4X 2020 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us 29 80 ...

Страница 30: ...age VISUALIZATION graphically displays the wavelength curve of the connected sensors The channels can be selected by clicking on the channel at Channel Selection The current wavelengths for all channels are displayed in the Current Values section 30 80 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us fos4X 2020 ...

Страница 31: ...ORD FIRMWARE UPDATE System Name Location Figure 14 System Name Location Table 29 System Name Location System name and location Section Content System Name Set name of the system System Location Set location of the system With a click on the button EDIT the editing for System Name and System Location can be activated fos4X 2020 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us 31 80 ...

Страница 32: ...sk Current subnet mask Default gateway Current default gateway SNTP Informations about SNTP Enabled Displays if SNTP is on Server IP Current set SNTP server address Syslog Informations about the system log Enabled Displays if the system log is on Server IP Current set server address for the system log With a click on the button EDIT the editing for the settings can be activated 32 80 MDE0835_4 10_...

Страница 33: ...splayed on the CAN MODE page In CAN mode With a click on the button SWITCH TO ETHERNET MODE the Blackbird OECD is rebooted and then works in Ethernet mode In Ethernet mode With a click on the button SWITCH TO CAN MODE the Blackbird OECD is rebooted and then works in CAN mode fos4X 2020 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us 33 80 ...

Страница 34: ...he information about the CAN bus See System State page 25 With a click on the button EDIT the editing for the following settings can be activated Bitrate CANopen Node ID Auto Operational Auto Recover From Bus OFF PDO Transmission Mode 34 80 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us fos4X 2020 ...

Страница 35: ...e page CAN SIGNAL SETTINGS displays the CAN settings for each channel See CANopen Objects page 40 With a click on the button EDIT the editing for the following settings can be activated Offset Scale Factor fos4X 2020 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us 35 80 ...

Страница 36: ...d Displays if a password is currently set for the settings Read access without password Displays if read access to the web interface is possible without password With a click on the button CHANGE OR DISABLE PASSWORD the editing for the password settings can be activated 36 80 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us fos4X 2020 ...

Страница 37: ...uality have the following meanings Signal Good Signal quality is good Measurements are possible without restrictions Signal Warning There are one or more warnings See Tab 32 Signal Quality page 37 Measurements are possible The measurements may be inaccurate due to boundary effects and environmental conditions Signal Problem There are one or more errors See Tab 32 Signal Quality page 37 Measurement...

Страница 38: ...hernet Interface In Ethernet mode the Blackbird OECD transmits the data in the form of UDP messages Communication takes place via IPv6 link local addresses The fos4X multicast address ff02 13c is used for device recognition For communication via the Ethernet interface fos4X recommends the use of the fos4X C API See fos4X C API documentation Time synchronization In Ethernet mode the Blackbird OECD ...

Страница 39: ...tion Part 1 Cabling and connector pin assignment CiA 303 2 Recommendation Part 2 Representation of SI units and prefixes CiA 303 3 Recommendation Part 3 Indicator specification For further informationen see www can cia org Byte order All CANopen messages are in the little endian byte order Node Guarding fos4X recommends Use Node Guarding to monitor the correct operation of the Blackbird OECD For m...

Страница 40: ...Reset 0x81 3 5 5 1 PDOs The Blackbird OECD transmits 2 PDOs PDO1 Table 34 Mapping in PDO1 0 1 2 3 4 5 6 7 2000sub1 Channel 1 2000sub2 Channel 2 2000sub3 Channel 3 2001sub1 Status 1 PDO2 Table 35 Mapping in PDO2 0 1 2 3 4 5 6 7 2000sub4 Channel 4 2000sub5 Channel 5 2000sub6 Channel 6 2001sub2 Status 2 3 5 5 2 CANopen Objects 0x1009 Manufacturer hardware version The object Manufacturer hardware vers...

Страница 41: ... ID 0x0000f4f4 Product code MDE0835 Revision number Software Version major version in upper 16 bit minor version in lower 16 bit Serial number 32 bit serial number of the measuring device 0x2000 Wavelength Data Description All values are 16 bit signed numbers Values between 32752 0x800F and 32768 0x8000 are error values See Tab 37 Error values in 0x2000 page 42 The other values are formed with the...

Страница 42: ...8002 No sensor detected on this channel 0x8003 Sensor temporarily unavailable for measurement light source regulates power 0x8004 Sensor or channel error 0x8005 0x800F Reserved 0x2001 Status The Blackbird OECD has 4 status fields Each 16 bit status field consists of 12 status bits and 4 counter bits 42 80 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us fos4X 2020 ...

Страница 43: ...ebserver Error The web server is not working The web interface is not available 5 Reserved Bit reserved for future assignment 6 Reserved Bit reserved for future assignment 7 Light Source Error A light source is not working 8 Temperature Error At least one component outside of its permissible operating temperature 9 Signal Low The signal from at least one sensor is low Measurements are still possib...

Страница 44: ...bit is incremented at each Sync frame 0x2001sub4 Status4 Table 41 Bits in 0x2001sub4 Status4 Bit Name Description 0 11 Reserved Bits reserved for future assignment 12 15 Sync counter Counter bits The counter bit is incremented at each Sync frame 0x2002 Lower Limit of the Measuring Range Description Object 0x2002 contains the lower limit of the measuring range for all channels Each sub object of 0x...

Страница 45: ...st channel Object 0x2003 is read only Sub Objects All sub objects are 32 bit unsigned integers The sub objects contain the 20 bit fix point representation of the wavelength To calculate the wavelength the raw value must be divided by 1048576 Sub Objects for channels that are not present contain the value 0 0x2004 Scale Factor Description Object 0x2004 contains the scale factor for all channels Eac...

Страница 46: ...he value for the first channel Object 0x2005 is read only Sub Objects All sub objects are 32 bit unsigned integers The sub objects contain the 20 bit fix point representation of the wavelength To calculate the wavelength the raw value must be divided by 1048576 Sub Objects for channels that are not present contain the value 0 46 80 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us fos4X 2020 ...

Страница 47: ...lackbird OECD is packed in a box with foam inserts One box contains up to 5Blackbird OECD The box has the dimensions 645 x 492 x 216 mm length x width x heigth 4 3 Unpacking Unpacking the the measuring device 1 Open the box 2 Remove the upper foam part from the box 3 Remove the measuring device from the box 4 Check the scope of delivery See Scope of Delivery page 47 5 If parts of the scope of deli...

Страница 48: ...g device in a different packaging 1 Choose a box so that the measuring device with padding material fits into the box 2 Open the box 3 Pack the measuring device in the box with padding material Make sure that the measuring device is protected against shocks from all sides 4 Close the box P Blackbird OECD ready for transport or shipping 4 5 Transport Transporting the Blackbird OECD u Handle the Bla...

Страница 49: ... Store the Blackbird OECD in a dry and clean place u Store the Blackbird OECD in a transport packaging See Packing page 48 u Observe the environmental conditions See Environmental Conditions page 15 fos4X 2020 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us 49 80 ...

Страница 50: ... the Control Cabinet Conditions Control cabinet with sufficient space for installation See Dimensions page 17 Control cabinet with permissible environmental conditions See Environmental Conditions page 15 Attaching the measuring device in the control cabinet u Attach the measuring device with 4x M6 screws and nuts TIP Control cabinets from fos4X provide 4 welding studs to attach the measuring devi...

Страница 51: ... 5 3 GF 3 81 connector to the CAN connector 5 4 Connecting the Power Supply TIP The measuring device has an integrated protection against reverse polarity If the polarity of the power supply is incorrect the measuring device is protected against damage Figure 23 PWR connector Connecting the power supply 1 Switch off the power supply 2 Attach the cable of the power supply to the MC 1 5 2 GF 3 81 co...

Страница 52: ...before connecting TIP The assignment of channels and sensors depends on the application of the Blackbird OECD If available Observe the color coding of the sensors and channels Material and equipment Table 42 Material and equipment Name Type Quantity GTIN Optical connector cleaner Recommended Fujikura OCC A Ultra 1 4251140910209 Connecting a Sensor Figure 24 E 2000 APC connector and socket 1 Lockin...

Страница 53: ...tive cover 3 on the connector 2 4 Clean the connector 2 with an optical connector cleaner See manual of the optical connector cleaner 5 Close the protective cover 3 6 Connect the connector 2 to the socket 4 P The connector 2 snaps into the socket 4 with a clicking sound P The sensor is connected 7 Repeat the procedure for all sensors fos4X 2020 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us ...

Страница 54: ...ace is displayed 2 Click on SETTINGS 3 fos4X recommends Set a password See Set or change the password page 54 4 fos4X recommends Set a system name and location See Settings page 31 5 Set a CANopen Node ID factory setting 3 6 Set the CANopen data rate to the data rate of the CAN bus factory setting 500 7 Set the network settings Web Interface page 23 6 3 Set or change the password Setting or changi...

Страница 55: ...Figure 25 No password set Figure 26 Password set 5 If no password is set Click on CREATE PASSWORD 6 If a password is set Click on CHANGE OR DISABLE PASSWORD Figure 27 Setting a password fos4X 2020 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us 55 80 ...

Страница 56: ...New Password 9 Enter the new password in the field at Verify Password 10 Optional Check the box at Read access without password The option Read access without password allows read access to the web interface without password 56 80 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us fos4X 2020 ...

Страница 57: ... page 59 7 2 Opening the Web Interface Opening the web interface 1 Connect the PC to the Ethernet network where the measuring device is located 2 Optional Using a RJ45 cable connect the PC to the ETH connector of the measuring device 3 Open a browser on the PC 4 Enter the IP Adresse of the measuring device in the address bar Factory setting 192 168 39 99 If the option READ ACCESS WITHOUT PASSWORD ...

Страница 58: ...k on LOGIN The page STARTING WEB APPLICATION is displayed Figure 29 Initialization of the Web Interface 6 Wait until the initialization is complete P The page HOME is displayed Figure 30 Home page 58 80 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us fos4X 2020 ...

Страница 59: ...ptions do not help to fix the problem u Contact fos4X Support See fos4X Support page 73 Table 43 Troubleshooting PWR LEDs Problem Possible cause Solution POWER LED on the PWR connector is not on Blackbird OECD not supplied with power u Check the power cable for damage u Check that the power cable is plugged in u Check that the power supply is switched on fos4X 2020 MDE0835_4 10_1 3 x_5 0_Blackbird...

Страница 60: ...use of the sensor 5 Open the page HARDWARE STATE 6 Check the temperature of the components 7 If the temperature of a component is too high Make sure that the temperature regulation in the control cabinet functions properly System not configured u Do a firmware update See Firmware Update page 70 STATUS LED on PWR connector does not light green See LEDs on the PWR connector page 18 System error 1 Ch...

Страница 61: ...he Blackbird OECD Turn the power supply off and on Signal Warning Wavelength Low or Wavelength High u Make sure that connected sensors are within the measuring range See Measurement and Sensors page 12 u Make sure that only permitted sensors are connected See Measurement and Sensors page 12 Temperature Too Low or Temperature Too High u Make sure to operate within specification See Environmental Co...

Страница 62: ... supply off and on Calibration Missing u Contact fos4X Support See fos4X Support page 73 No Signal u Make sure that a sensor is connected u Clean the optical connectors See Cleaning E 2000 APC Connections page 68 u Reboot the Blackbird OECD Turn the power supply off and on Device Initializing u Wait until the initialization is complete 8 3 Troubleshooting Ethernet TIP If the described options do n...

Страница 63: ...ontact fos4X Support See fos4X Support page 73 8 4 1 Troubleshooting CAN Connection and LEDs Table 46 Troubleshooting CAN Connection and LEDs Problem Possible cause Solution Cable connection defective u Check the CAN cable for damage Replace damaged cables u Make sure that the CAN cable is connected correctly CAN Interface not configured correctly u Configure the CAN interface See Commissioning th...

Страница 64: ...igured correctly See Commissioning the CAN Interface page 54 u Make sure that the Bus Master is configured correctly u Make sure that the bit rate of the CAN bus is configured correctly u Check the CAN cable for damage Replace damaged cables u Make sure that the CAN cable is connected correctly u Do a firmware update See Firmware Update page 70 8 4 2 Troubleshooting CAN Object 0x2000 TIP The numbe...

Страница 65: ...ng E 2000 APC Connections page 68 u Reboot the Blackbird OECD Turn the power supply off and on u Check the signal quality See Troubleshooting Signal Quality page 61 Object 0x2000 has value 0x8003 Sensor temporarily unavailable for measurement light source regulates power u Wait until regulation is complete u Reboot the Blackbird OECD Turn the power supply off and on u Check the signal quality See ...

Страница 66: ...is u Wait until the automatic troubleshooting is complete u Reboot the Blackbird OECD Turn the power supply off and on Bit 4 of 0x2001sub1 set See 0x2001 Status page 42 Webserver Error u Reboot the Blackbird OECD Turn the power supply off and on Bit 7 of 0x2001sub1 set See 0x2001 Status page 42 Light Source Error u Reboot the Blackbird OECD Turn the power supply off and on Bit 8 or 11 of 0x2001sub...

Страница 67: ...000 APC Connections page 68 u Reboot the Blackbird OECD Turn the power supply off and on u Check the signal quality See Troubleshooting Signal Quality page 61 Bit 10 of 0x2001sub1 set See 0x2001 Status page 42 Warning u Check which warnings are pending See Opening the Web Interface page 57 fos4X 2020 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us 67 80 ...

Страница 68: ...ance 9 2 1 Cleaning E 2000 APC Connections Figure 31 E 2000 APC connector and socket 1 Locking mechanism 2 Connector 3 Protective cover 4 Socket 5 Dust cap Table 49 Material and equipment Name Type Quantity GTIN Optical connector cleaner Recommended Fujikura OCC A Ultra 1 4251140910209 Cleaning E 2000 APC connections 1 Release the locking mechanism 1 of the connector 2 2 Remove the connector 2 fro...

Страница 69: ...000 APC connection is cleaned 10 Repeat the procedure for all connections 9 2 2 Replacing the PWR Connector Table 50 Material and equipment Name Type Quantity GTIN Power supply connector MC 1 5 2 GF 3 81 1 4017918050313 Figure 32 PWR connector Replacing the PWR connector 1 Switch off the power supply 2 Remove the MC 1 5 2 GF 3 81 connector from the PWR connector 3 Remove the power supply cable fro...

Страница 70: ...AN connector 3 Remove the CAN cable from the MC 1 5 3 GF 3 81 connector 4 Dispose of old MC 1 5 3 GF 3 81 connector in accordance with local regulations 5 Attach the CAN cable to the MC 1 5 3 GF 3 81 connector Observe the assignment of the pins High H Low L and Ground G at the CAN connector 6 Connect the MC 1 5 3 GF 3 81 connector to the CAN connector 7 Switch on the power supply 9 2 4 Firmware Up...

Страница 71: ...mplete 9 Reboot the Blackbird OECD Turn the power supply off and on P Firmware update complete 9 2 4 2 Firmware Update via fos4X Firmware Flasher Table 52 Material and equipment Name Type Quantity GTIN PC Ethernet connector RJ45 1 Ethernet cable RJ45 1 Software fos4X Firmware Flasher f3 exe 1 Conditions Firmware archive for the Blackbird OECD available IPv6 protocol enabled on the PC Doing a firmw...

Страница 72: ... OECD is not displayed a Make sure that the PC is connected to the ETH port of the Blackbird OECD b Make sure that the correct network adapter is selected c Make sure that the IPv6 protocol is enabled on the PC d Make sure that the PC has a valid IPv6 link local address e Click on in Network Interface 7 Click on in Firmware Archive 8 Select the firmware archive 9 Click on OPEN The firmware archive...

Страница 73: ...RMWARE b Only if only FLASH BOOTLOADER FIRMWARE is displayed with a green check mark Click on FLASH BOOTLOADER FIRMWARE The status of the update is displayed in a new window 13 Wait until the firmware update is complete 14 Disconnect the PC from the Blackbird OECD 15 If necessary Connect the cable to the ETH connector 16 Reboot the Blackbird OECD Turn the power supply off and on P Firmware update ...

Страница 74: ...turer GTIN Power supply connector MC 1 5 2 GF 3 81 Phoenix Contact 4017918050313 CAN connector MC 1 5 3 GF 3 81 Phoenix Contact 4251140902143 Accessories Table 54 Accessories Name Type Manufacturer GTIN Optical connector cleaner OCC A Ultra Fujikura 4251140910209 74 80 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us fos4X 2020 ...

Страница 75: ... Connector 3 Protective cover 4 Socket 5 Dust cap 2 Disconnect the optical connections a Release the locking mechanism 1 of the connector 2 b Remove the connector 2 from the socket 4 c Attach a dust cap 5 on the socket 4 d Repeat the procedure for all sensors 3 Remove the power cable from the PWR connector 4 Remove the RJ45 cable from the ETH connector 5 Remove the CAN cable from the CAN connector...

Страница 76: ...assembled by fos4X or a specialized disposal company 10 3 Disposal NOTICE Risk to the environment due to incorrect disposal of the product Improper disposal of the product may cause damage to the environment u Observe applicable environmental regulations The Blackbird OECD is an electrical appliance and must not be disposed of with household waste Disposing of the Blackbird OECD u Contact fos4X Su...

Страница 77: ... the variant 30 Fig 14 System Name Location 31 Fig 15 Network Settings 32 Fig 16 CAN Mode 33 Fig 17 CAN Bus Settings 34 Fig 18 CAN Signal Settings 35 Fig 19 System Password 36 Fig 20 Firmware Update 37 Fig 21 ETH connector 50 Fig 22 CAN connector 51 Fig 23 PWR connector 51 Fig 24 E 2000 APC connector and socket 52 Fig 25 No password set 55 Fig 26 Password set 55 Fig 27 Setting a password 55 Fig 28...

Страница 78: ...AN Connector 22 Tab 22 Factory Settings 22 Tab 23 Home Information on the home page 24 Tab 24 CAN State Informations about the CAN interface 26 Tab 25 Signal State Informations about each channel 26 Tab 26 CAN State Informations about the CAN interface 27 Tab 27 FBG Channel States Informations about each channel 28 Tab 28 Light Source States Informations about the light source 28 Tab 29 System Nam...

Страница 79: ...shooting CAN Connection and LEDs 63 Tab 47 Troubleshooting CAN Object 0x2000 64 Tab 48 Troubleshooting CAN Sub Object 0x2001Sub1 66 Tab 49 Material and equipment 52 Tab 50 Material and equipment 69 Tab 51 Material and equipment 70 Tab 52 Material and equipment 71 Tab 53 Spare Parts 74 Tab 54 Accessories 74 fos4X 2020 MDE0835_4 10_1 3 x_5 0_Blackbird OECD_ 6CH CNO en us 79 80 ...

Страница 80: ......

Отзывы: