![Fora TD-3261G Скачать руководство пользователя страница 55](http://html1.mh-extra.com/html/fora/td-3261g/td-3261g_owners-manual_2316825055.webp)
55
3. If the control solution testing result is out of range:
POSSIBLE CAUSE
WHAT TO DO
Error in performing the test.
Read instructions thoroughly and
repeat the test again.
Control solution vial was poorly
shaken.
Shake the control solution
vigorously and repeat the test
again.
Expired or contaminated control
solution.
Check the expiration date of the
control solution.
Control solution that is too warm or
too cold.
Control solution, monitor, and test
strips should be at room
temperature (20
°C
to 25
°C
/ 68
°F
to 77
°F
) before testing.
Defective test strip.
Repeat the test with a new test
strip.
Monitor malfunction.
Please contact customer service.
Improper working of meter and
test strip.
Please contact customer services.
Содержание TD-3261G
Страница 1: ...Owner s Manual FORA TD 3261G Blood Glucose Plus Blood Pressure Monitoring System...
Страница 2: ...2...
Страница 10: ...10 Monitor Overview Front Side Back Pressure Cuff 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18...
Страница 20: ...20 Step 2 Press to switch between General AC PC and QC mode...
Страница 70: ......