Important Control Solution Information
• Use only our control solutions with your monitor.
• Do not use the control solution beyond the expiry date or 3 months
after first opening. Write the opening date on the control solution
vial and discard the remaining solution after 3 months.
• It is recommended that the control solution test should be done at
room temperature (20°C to 25°C / 68°F to 77°F). Make sure your
control solution, monitor, and test strips are at this specified
temperature range before testing.
• Shake the vial before use, discard the first drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an
accurate result.
• Store the control solution tightly closed at temperatures between
2°C and 30°C (36°F and 86°F). Do NOT freeze.