![Fora TD-3261G Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/fora/td-3261g/td-3261g_owners-manual_2316825023.webp)
23
Step 4
Read and Compare the Result
After counting down to 0, the test result of
control solution will appear on the display.
Compare this result with the range printed
on the test strip vial and it should fall
within this range. If not, please read the
instructions again and repeat the control
solution test.
109~165
Normal
Out-of-range results
If you continue to have test results fall outside the range printed on
the test strip vial, the monitor and strips may not be working properly.
Do NOT
test your blood. Contact the local customer service or place of
purchase for help.
NOTE
• The control solution range printed on the test strip vial is for control
solution use only. It is not a recommended range for your blood glucose
level.
• See the
MAINTENANCE
section for important information about your
control solutions.
Содержание TD-3261G
Страница 1: ...Owner s Manual FORA TD 3261G Blood Glucose Plus Blood Pressure Monitoring System...
Страница 2: ...2...
Страница 10: ...10 Monitor Overview Front Side Back Pressure Cuff 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18...
Страница 20: ...20 Step 2 Press to switch between General AC PC and QC mode...
Страница 70: ......