EN-5
Important!
If the cuff is relatively lower (higher) than the heart, the
obtained blood pressure value could be higher (lower) than
the actual value. A 15 cm difference in height may result in an
error around 10 mmHg.
4. Press
the
Power
button.
The device will turn on and the cuff will begin to inflate
automatically. Remain still and do not talk or move during the
measurement.
Measuring Blood Pressure
This device provides you different ways to measure your blood
pressure.
When the device detects an irregular heart beat, the IRB symbol
is shown as “
” instead of “
”. If the problem still persists,
please consult your doctor.
Important!
Always apply the pressure cuff before turning on the device.
A. Single Measurement
Perform an individual blood pressure measurement.
1. Press
the
Power
button. All the LCD symbols will appear. Then
the cuff will begin to inflate automatically.
Note:
• When the device is connected to your mobile device (via
Bluetooth),you can also initiate the measurement by tapping
.
• The cuff will begin inflate automatically.
2. The heart symbol “
” will flash when a pulse is detected
during the inflation.