9. PROGRAMMING
9.1. PROGRAMMING PROCEDURE
Default kV and mA values for each program are assigned by default in factory
Such values can be re-loaded using a dedicated service function.
If you want to modify these default values please follow the instruction below.
By keeping pressed the BACK key for 5 s the system enters Service Mode
Service function can be selected using
the INCREASE or DECRESE keys on the RIGHT side of the display.
Scroll the menu until Exposure Settings is displayed
Press the PROGRAM key to enter Exposure Settings.
NOTE.
Keep pressed the TEST key to reset all values to default (factory) conditions
Select the exposure program by pressing the program key:
Select the patient build by pressing the PATIENT key,
Small, Medium, Large, Extra Large
Change the kV level to the desired value by pressing the
INCREASE or DECRESE keys at the LEFT side of the display
Change the mA level to the desired value by pressing the
INCREASE or DECRESE keys at the RIGHT side of the display
Keep pressed the LIGHT key to store the selected value
Repeat kV setting and mA setting
for each patient build (if requested) and for each exposure program.
Press BACK key to exit EXPOSURE SETTINGS
Press TEST Mode key to exit SERVICE MODE
Do verify proper setting by changing PATIENT build for each exposure program