【 4.1 】 Switching Of Signal Types
Press 【 FN 】 on the main interface.After the 【 FN 】 button turns red, if you want to
switch the input signal type, press the 【 IN 】 button to enter the input setting
If you want to switch the output signal type, press 【 OUT 】 key to enter the output
setting.After entering the setting interface, you can directly select the desired signal
type from left to right.Press 【 FN 】 to exit to the main interface
There are also a variety of signal modes available for switching under various signal
types.Various signal types also have different parameters that can be set
【 4.2 】 mA Current Signal
【 Mode Switch 】
: When the main interface 【 FN 】 is white, press 【 IN/OUT 】 to
switch the display of input/output actual value and engineering quantity
【 Output Open Circuit Detection 】
: When the output is turned on, if the interface
displays open circuit and flashes, it means that the output signal is not connected
or the output is close to zero
【 Output Circuit Voltage 】
:FNIRSI SG-004A can set current output no-load
voltage.The purpose of 12V is to save power.Some devices require a 24V drive
capability
【 Output Range 】
:The limit range of the current output can be set
【 4.3 】 V Voltage Signal
【 Mode Switch 】
: When the main interface 【 FN 】 is white.Press 【 IN/OUT 】 to switch
the display of actual value and engineering quantity of input/output
【 Overcurrent Protection 】
: If the output voltage is shorted, the output will
automatically shut down
【 Output Range 】
:The limit range of the voltage output can be set
�、DISPLAY & SIGNAL DESCRIPTION
��
PAGE
【 Input Signal Settings 】
【 Output Signal Settings 】