【 Parameter Settings 】

:

(1) 

Input upper limit/input lower limit: the amplitude range of the input signal

(2)

 Output upper limit/output lower limit: the amplitude range of the output signal

(3)

 Overrun allowed: For example, 1-5V is converted into 4-20ma.If the user checks 

this option, when the input is oV, the output will follow to OmA.If the user does not 

check it, then the output is 4mA and the output is locked within the set range

(4)

 Automatic operation:After checking this option, the signal conversion will be 

automatically started at the next power-on, and the signal conversion will be 

automatically canceled when the signal conversion is exited

【 Instructions 】

:

【 Start and Exit 】:When the main interface 【 FN 】 is red, press the key 【 CONVRET 】 

to start/exit the signal conversion mode.After startup, the backlight of the up, 

down, left and right keys will turn off and have no function

【 5.4 】  Program Output

The programming output can automatically complete N cycles of signal output 

according to the parameters set by the user.This function is mainly used for the aging 

test of electric valve or pneumatic valve,Or test work such as PLC program debugging

【 Parameter Settings 】

:

(1)

Mode: It is divided into three modes: single rise, single fall and cycle

(2)

Number of loops: 0 is infinite, 30000 is the maximum

(3)

Start value: The value at each start

(4)

End value: The value at the end of each

(5)

Increment step value: The value of each increment

(6)

Increment step time: The time between each increment

(7)

End stop: The time delay at the end value

(8)

Decrement step value: The value to decrease each time

(9)

Decrease step time: The time between each decrease

(10)

Loop stop: Time delay at start value

【Supplementary Parameter Description】

:

The start value and end value of single rise and single fall are selected according to 

high and low,and does not automatically close

【 Instructions 】

:

(1)

Start and Exit:When the main interface 【 FN 】 is red, press the left key 【 AUTO 】 

to start/exit the programming output mode

(2)

Up, Down, Left and Right keys: After starting the programming output mode, the 

backlight of the up, down, left and right keys will turn blue.The functions of the up, 

down, left and right buttons will also change, that is, the top is a single run, the bottom 

is a program reset, the left is a single-step run, and the right is a Stop/Continue

��

PAGE

Содержание SG-004A

Страница 1: ...Multi functional Signal Generator OPERATING MANUAL SG 004A...

Страница 2: ...6 mV Millivolt Signal 07 08 4 8 Ohm Resistance Signal 4 Display Signal Description 2 1 Speci cations 01 2 3 Interface Terminal 02 2 2 Function Introduction 02 2 4 Power Charge 03 2 Product Description...

Страница 3: ...ts 0 01mV 1 1 1 1 1 1 1 1 1 1 0 1 0 1 0 1 100 500K 100 100 1MK 2K 2K 2K 2K 2K 2K 2K 2K 2K 2K 2K Signal Range Precision Resolution Impedance PAGE Plug the meter into the PC and turn it on Wait for the...

Страница 4: ...monotonous rise monotonous fall and cycle Real time Curve The change trend of input and output signals can be displayed by drawing a curve and the sampling time of the curve can be set It can also aut...

Страница 5: ...ode Button green light Preset output mode Button blue light Programming mode No button light Conversion mode BUTTON PANEL INTRODUCTION Button Function Operation Description RUN Light The power on on o...

Страница 6: ...output the down key reset the left key end the output the right key pause continue ON OFF con rmation button when the main interface turn on o the output When the FN is red light click to turn on o t...

Страница 7: ...e display of input output actual value and engineering quantity Output Open Circuit Detection When the output is turned on if the interface displays open circuit and ashes it means that the output sig...

Страница 8: ...ted The output frequency range can be switched by pressing OUT when the main interface FN is white Operating Instructions In PWM Mode PWM displays two parameters the frequency value that is the period...

Страница 9: ...etection type of the input signal by the IN port In this case select mA for the input signal type Then the IN port can be combined with the IN port to form a 24V output And at the same time detect the...

Страница 10: ...e of input output EXTENSIONS 5 1 Engineering When the signal value of analog quantity is used in the industrial eld it is generally necessary to convert it into engineering quantity value Therefore in...

Страница 11: ...g test of electric valve or pneumatic valve Or test work such as PLC program debugging Parameter Settings 1 Mode It is divided into three modes single rise single fall and cycle 2 Number of loops 0 is...

Страница 12: ...the top of the screen will ash to display the record number View Record Select the record you want to view in the setting interface history record ON OFF con rm to enter the view When entering the vie...

Страница 13: ...sed to 24V as the negative pole of the input signal the input and output are performed at the same time WIRING INSTRUCTIONS PLC Process instrument etc All switched by electronic switches All round Sig...

Страница 14: ...indicated by the technical indicator It may cause the instrument to work abnormally or even damage Error in input and output It is normal that there may be errors in input and output Because the inpu...

Страница 15: ...name Multi functional Signal Generator Brand Model FNIRSI SG 004A Manufacturer Shenzhen FNIRSI Technology Co Ltd Website www fnirsi cn Address Building C Weihuada Industrial Park Dalang Street Longhu...

Страница 16: ...e 03 is frequency 04 is millivolt 05 is resistance The upper four bits are the type 1 8 represent S B E K R J T N in turn Lower four bit mode 1 is mV 2 is thermocouple 3 is WR thermocouple Enter Sched...

Страница 17: ...Download Manuals and Firmware Manufacturer Shenzhen FNIRSI Technology Co Ltd Website www fnirsi cn Address Building C Weihuada Industrial Park Dalang Street Longhua District Shenzhen Guangdong China...

Отзывы: