Operating manual EMGZ307
12
Gain and offset with fixed setting
If the signal is processed by a control unit
(for ex. a PLC), configuration with fixed
gain and offset is recommended (offset =
0; gain = 1111). Therefore, the jumpers
have to be set according to fig. 8. With this
configuration, the good thermal
characteristics of the hybrid module are
best utilized. The output signal will be
10V and 20mA when the nominal force of
the sensor is applied.
7.4
Configuring the Outputs
The measuring amplifier provides a tension output (±10V) and a current output
(0...20mA or 4...20mA) which are active simultaneously. The signal of the current
output is configured as follows:
Current output 0...20mA
The jumpers have to be set according to
fig. 9a. This is the standard FMS factory
setting.
Current output 4...20mA
The jumpers have to be set according to
fig. 9b.
fig. 8:
Jumper setting for fixed gain
and offset
E307006eb
fig. 9a:
Jumper setting for current
output 0...20mA
E307006ec
fig. 9b:
Jumper setting for current
output 4...20mA
E307006ed