Albuterol
sulfate
Ipratropium
bromide
Cromolyn
sodium
Coarse particle dose (μ)
69.2 +/- 36.1
7.4 +/- 3.8
85.7 +/- 37.2
Fine particle dose (μ)
969.7 +/- 74.2
223.4 +/- 16.3
872.7 +/- 80.9
Extra fine particle dose (μ)
374.9 +/- 42.0
89.8 +/- 6.7
356.7 +/- 47.4
Coarse fraction (>4.7μ)
6.7%
3.1%
8.9%
Fine fraction (<4.7μ)
92.4%
96.3%
89.7%
Extra fine fraction (<1.0μ)
35.7%
38.7%
36.6%
All data shown at a 95% confidence level. Medications were tested 3
times each on 3 devices for a total of 27 sample data points.
Aerosol Performance
Medical Electrical Equipment
needs special precautions
regarding EMC and needs
to be installed and put into
service according to the EMC
information provided in this
User Guide.
Portable and Mobile RF
Communications Equipment
can affect Medical Electrical
Equipment.
The use of accessories,
transducers and/or cables other
than those specified, with the
exception of those sold by the
manufacturer as replacement
ELECTROMAGNETIC
COMPATIBILITY
With the increased number
of electronic devices, such
as computers and cellular
phones, medical devices in
use may be susceptible to
electromagnetic interference.
This interference may result
in incorrect operation of the
medical device and create a
potentially unsafe situation.
Do not use cellular phones and
other electronic devices that
generate electromagnetic
fields near the medical device.