Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment Guidance
Power
Frequency
50/60Hz
Magnetic Field:
IEC61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment – Guidance
Voltage
dips and
interruptions:
IEC61000-
4-11
0 % UT: 0.5 cycle
at 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°.
0% UT: 1 cycle
70% UT: 25 cycles for
50 Hz
Single phase: at 0°
0 % UT; 250/300 cycle
for 50 Hz and 60 Hz
0 % UT: 0.5 cycle
at 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°.
0% UT: 1 cycle
70% UT: 25 cycles for
50 Hz
Single phase: at 0°
0 % UT; 250/300 cycle
for 50 Hz and 60 Hz
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the FLYP
pe1200m requires
continued charging
during mains power
interruptions, it is
recommended that
the FLYP pe1200m
be powered from an
uninterruptible power
supply or battery.
Guidance and Manufacturer’s Declaration – Immunity
The FLYP™ pel200m is intended for use in the electromagnetic environment
specified below. The customer or user of the FLYP pel200m should ensure
that it is used in such an environment.