3. Let the ethyl alcohol sit for 10
minutes. Then shake off the ethyl
alcohol from the Nebulizer Disk.
4. Turn FLYP over. Open the
Reservoir Cover and remove the
Medication Reservoir.
5. Repeat steps 2 and 3 on this side
of the Nebulizer Disk.
6. Place a few drops of distilled
water on the Nebulizer Disk.
Then shake off the distilled
water. Repeat on the other side
of the Nebulizer Disk.
Then remove the ethyl alcohol
from the interior:
1. Reinsert the Medication
Reservoir and Mouthpiece.
2. Fill the Medication Reservoir with
distilled water.
3. Plug the Stopper.
4. Close the Reservoir Cover.
5. Gently raise the Mouthpiece
with your index finger.
6. Turn FLYP™ on and allow the
mist to flow from the raised
Mouthpiece for 10 minutes.
7. Unplug the Stopper and remove
the Medication Reservoir.
8. Shake excess distilled water from
the Medication Reservoir.
9. Let the Medication Reservoir dry
on a clean, dry towel.
10. Let the Main Unit dry on a
clean, dry towel.
11. Allow the Main Unit to dry
fully before using it again.
Disinfecting the Main
Unit’s exterior:
Before disinfecting the Main Unit’s
exterior, make sure the USB Cover
is closed tightly. Disinfect the Main
Unit’s exterior by wiping it with a
clean, dry towel lightly moistened
with ethyl alcohol. Then allow
the Main Unit to sit for 5 minutes.
Moisten a clean, dry towel with
water and wipe the Main Unit to
remove the ethyl alcohol. Allow to
air dry.
NEVER immerse the Main Unit or
USB Wall Charger in ethyl alcohol.
The lithium-ion battery inside the
Main Unit may be dangerous
when wet.
NEVER use window cleaners,
aerosol sprays, solvents, ammonia,
or abrasives to disinfect FLYP.