TROUBLESHOOTING
Most problems with the device
can be solved by following this
advice.
The following table describes
possible troubles, their causes,
and what corrective action to
take.
Section 7: TROUBLESHOOTING
If these corrective actions do
not return FLYP™ to full working
order, read the next section,
“Support.”
TROUBLE
POSSIBLE CAUSE
CORRECTIVE ACTION
The
nebulization
rate is
extremely low.
The battery is low.
Charge FLYP using a USB
port. Refer to pages 22-23.
An air bubble
has formed in the
medication’s pathway.
Tap the back of the Main
Unit with your index finger
until normal flow resumes.
The Nebulizer Disk has
become clogged with
medication.
Nebulize distilled water for
1 minute. Refer to
pages 29-32.
The On-Off
Indicator
Light is on,
but there is no
mist.
The Medication
Reservoir is not correctly
inserted.
Remove the Medication
Reservoir and install it
correctly.
The Medication
Reservoir is not filled.
Fill with medication. Refer to
pages 24-26.
Check that FLYP has
been properly cleaned.
Nebulize distilled water for
1 minute. Refer to pages 29-32.