5.6.3
Error list
A protocol of the errors that have occurred is shown in this window. A signal tone sounds when an
error occurs. The tone can be turned off by touching the display. Press the “REVERSE” button to
enter the list and page up and down with the “Fcs. Up” and “Fcs. Dwn” buttons. Close the window
with the “X” in the upper right corner.
Message:
Solution:
Motor Error
Problem with the autosampler motor, Initialize AS, contact the
service
AS position wrong
Initialize AS
AS not in furnace position
Initialize AS
Door open
Close the door, it must always be closed!
Cooling tome > melting time
Melting time in the application is too short, lengthen them
AS X‐axis error
AS error, contact the service
There is no mould on the holder No casting dish
Remove holder
After an emergency stop, there is still a holder on the AS, it
must be manually removed. WARNING HOT!
Init X error
AS error, contact the service
Init Y error
AS error, contact the service
Close furnace
Furnace is open, close manually
Error I47
AS error, contact the service
Error I46
AS error, contact the service