5.6.2
Autosampler
Reset Autosampler
Must be conducted when the error message “AS position
wrong” occurs to bring the autosampler to the zero position.
Unload sample from furnace
Function to remove a sample from the furnace after an
emergency stop. At the end, the autosampler stops behind
station 1. The sample must then be removed manually
(WARNING: hot!) The autosampler must be returned to the
zero position with the Reset autosampler button.
Open Furn.
The furnace is opened completely; allowing a view of the
inside for control. The furnace closes automatically after 5
seconds.
Close Furnace
Closes the furnace manually after an emergency stop.