General Specifications
Maximum between any Terminal
and Earth Ground........................ 1000 V ac rms, continuous, 1500 V dc
(Single-range)
Input Impedance ......................... 10 M
Ω
/ <10 pF
Dimensions ................................. 138 mm x 75 mm x 45 mm
Input terminal .............................. 4 mm banana jacks
Output terminal............................ BNC (female)
Weight ......................................... 295 g, batteries included
Accessory Cable ......................... BNC to banana male to male, 1.6 m
Temperature
Operating ................................. -10 °C to 50 °C
Storage .................................... -30 °C to 60 °C
Operating Humidity ..................... Non-condensing (
≤
10 °C)
≤
90 % RH (at 10 °C to 30 °C)
≤
75 % RH (at 30 °C to 40 °C)
≤
45 % RH (at 40 °C to 50 °C)
Altitude
Operating ................................. 2000 m
Storage .................................... 12 000 m
Ingress Protection ....................... IEC 60529: IP40 (Terminals Mated)
Battery
Type ......................................... 4 AA, IEC LR6
Battery Life............................... 168 hours (typical)
Safety
General .................................... IEC 61010-1: Pollution Degree 2
Measurement ........................... IEC 61010-2-030: CAT III 1000 V
CAT IV 600 V DC 1500 V
Electromagnetic Compatibility (EMC)
International ............................. IEC 61326-1: Portable
Electromagnetic Environment
IEC 61326-2-2
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or use
conductively coupled radio-frequency energy which is necessary for
the internal functioning of the equipment itself.
Class A: Equipment is suitable for use in all establishments other
than domestic and those directly connected to a low voltage power
supply network which supplies buildings used for domestic purposes.
There may be potential difficulties in ensuring electromagnetic
compatibility in other environments, due to conducted and radiated
disturbances.
Caution: This equipment is not intended for use in residential
environments and may not provide adequate protection to radio
reception in such environments.
Emissions that exceed the levels required by CISPR 11 can occur
when the equipment is connected to a test object.