Battery

XW

 Warning

To prevent possible electrical shock, fire, or personal injury:

• Disconnect the Product from any measurement device 

before you install or replace the batteries.

• Replace the batteries when the low battery indicator 

shows to prevent incorrect measurements.

To install the batteries, see Figure 

1

:

1.

Disconnect all Probe connections.

2.

Make sure that the power switch is OFF.

3.

Use a flat-blade screwdriver to rotate the battery door fastener to 
the unlocked position.

4.

Remove the battery door.

5.

Install the new batteries with the correct polarity.

6.

Lock the battery door.

Table 2. The Product

Item

Description

Battery indicator (red flash = low battery, green flash = 
battery ok)

Power switch

Output terminal

Battery door fastener

Input terminals

Output Test lead

1

2

3

4

U1500S

HIGH 

VO

LTAG

E DIF

FER

ENTIA

L PR

OBE

6

5

Содержание U1500s

Страница 1: ...ngapore 65 6799 5566 Anywhere in the world 1 425 446 5500 Customer Service Center Fluke China 400 810 3435 Maintenance Service Center Fluke China 400 921 0835 To view print or download the latest manu...

Страница 2: ...the battery door is opened Do not work alone Do not exceed the Measurement Category CAT rating of the lowest rated individual component of a Product probe or accessory Do not touch voltages 30 V ac rm...

Страница 3: ...off the Product after test completion then unplug the test leads Conforms to relevant Australian Safety and EMC standards Conforms to relevant South Korean EMC Standards Measurement Category III is ap...

Страница 4: ...Disconnect all Probe connections 2 Make sure that the power switch is OFF 3 Use a flat blade screwdriver to rotate the battery door fastener to the unlocked position 4 Remove the battery door 5 Insta...

Страница 5: ...Have an approved technician repair the Product If the Product does not perform properly use these steps to help isolate the problem 1 Verify that battery capacity meets the requirements 2 Verify that...

Страница 6: ...V DC 1500 V Electromagnetic Compatibility EMC International IEC 61326 1 Portable Electromagnetic Environment IEC 61326 2 2 CISPR 11 Group 1 Class A Group 1 Equipment has intentionally generated and or...

Страница 7: ...R 15 subpart B This product is considered an exempt device per clause 15 103 Electrical Specifications Measurement Range 1000 V ac rms 1500 V dc Accuracy 0 1 of reading 0 1 of Range Bandwidth 100 kHz...

Страница 8: ...on then send the product to that Service Center with a description of the problem THIS WARRANTY IS YOUR ONLY REMEDY NO OTHER WARRANTIES SUCH AS FITNESS FOR A PARTICULAR PURPOSE ARE EXPRESSED OR IMPLIE...

Отзывы: