29
Maintenance
Maintenance
W
Warning
X
These servicing instructions are for use by
qualified personnel only. To avoid electric
shock, do not perform any servicing other than
that contained in the operating instructions
unless you are qualified to do so.
Disconnect clips from any metallic connections
before performing any maintenance.
W
Caution
Do not use CRC Cable Clean
®
or any similar
chlorinated solvent on the tester. Doing so will
damage the tester.
There are no user serviceable components or adjustments in
the tester. Do not open the housing as handling of the PC
board could remove the moisture protection coating or
apply a static charge that will damage sensitive components.
Note
Opening the housing will void the warranty.
Moisture will not harm the tester. However, moisture can
provide a leakage path that may conduct hazardous
voltages to you. DO NOT USE the tester if wet.
If moisture should get inside the tester, let the tester dry at
normal room temperature for 24 hours. DO NOT HEAT THE
tester.
The tester may be cleaned by using a soft cloth with soap
and water. Do not use a petroleum-based or chlorinated