Multifunction PV Analyzer
Specifications
3
Vibration
............................................................. MIL-PRF-28800F: Class 2
Ingress Protection
.......................................... IEC 60529:IP40
Computer Interface
........................................ IR (serial) and Bluetooth
Wireless connection compatibility
.......... IRR2-BT
Accuracy
The accuracy specification is defined as ± (% r digit counts) at 23
o
C ± 5
o
C,
≤
80 % RH.
The accuracy specification for 0
o
C to 18
o
C and 28
o
C to 50
o
C: 0.1 x (accuracy specification)
for each °C.
Protective Conductor Resistance R
LO
PV Model/PV String, Open-Circuit Voltage, (V
OC
)
PV Model/PV String, Short-Circuit Current, (I
S/C
)
Display Range
Measurement Range
Resolution
Accuracy
0.00
Ω
to 19.99
Ω
0.20
Ω
to 19.99
Ω
0.01
Ω
± (2 % + 2 Digit)
20.0
Ω
to 199.9
Ω
20.0
Ω
to 199.9
Ω
0.1
Ω
± (2 % + 2 Digit)
200
Ω
to 2000
Ω
200
Ω
to 2000
Ω
1
Ω
± (5 % + 2 Digit)
Test current
≥
200 mA (
≤
2
Ω
+ R
COMP
)
[1]
Test voltage
4 V
DC
to 10 V
DC
Polarity reversing
Yes
Test lead zero (Rcomp)
up to 3
Ω
Live circuit detection
Inhibits test if terminal voltage >50 V ac/dc (typical) detected prior to initiation
of test.
[1] The number of 200 mA @ 0.1
Ω
continuity tests with a set of new batteries is >1000.
Display Range
Measurement Range
Resolution
Accuracy
0.0 V to 99.9 V
5.0 V to 99.9 V
0.1 V
± (0.5 % + 2 Digit)
100 V to 1000 V
100 V to 1000 V
1 V
Polarity test
Yes
Live circuit detection
Inhibits test if terminal voltage >5 V ac detected prior to initiation of test.
Display Range
Measurement Range
Resolution
Accuracy
0.0 A to 20.0 A
0.2 A to 20.0 A
0.1 A
± (1 % + 2 Digit)
Live circuit detection
Inhibits test if terminal voltage >5 V ac (typical) detected prior to initiation of test.