Calibration Adjustment
4.6 Final Calibration
4
4-9
Table 4-2. HF Gain Calibration Points Slow
Cal step
5500A Setting
(1 kHz, MODE
wavegen,
WAVE square)
Test Tool Input Signal
Requirements
(1 kHz square, t
rise
<2
μ
s,
flatness after rising edge:
<0.5% after 4
μ
s)
HF-Gain AB (CL 0609)
25V
25V
HF-Gain A (CL 0612),
HF-Gain B (CL 0632)
HF-Gain A (CL 0615), HF-Gain B (CL 0635)]
1)
50V 50V
1)
After starting the first step in this table cell, these steps are done automatically.
4.6.2 Delta T Gain, Trigger Delay Time & Pulse Adjust Input A
Proceed as follows to do the calibrations:
1.
Press
to select calibration step
Delta T (CL 0700):IDLE
2.
Connect the test tool to the 5500A as shown in Figure 4-4.
ST8004.WMF
Figure 4-4. 5500A Scope Output to Input A
3.
Set the 5500A to source a 1V, 1 MHz fast rising (rise time
≤
1 ns) square wave
(SCOPE output, MODE edge).
4.
Set the 5500A to operate (OPR).
5.
Press
to start the calibration.
The Delta T gain, Trigger Delay (CL0720), and Pulse Adjust Input A (CL0640) will
be calibrated.
6.
Wait until the display shows
Pulse Adj A (CL 0640):READY
.
7.
When you are finished, set the 5500A to Standby.
8.
Continue at Section 4.6.3.
Содержание ScopeMeter 123
Страница 6: ......
Страница 12: ......
Страница 16: ......
Страница 28: ...Fluke 123 124 125 Service Manual 2 14 Figure 2 2 Max Input Voltage v s Frequency for VP40 10 1 Voltage Probe...
Страница 30: ...Fluke 123 124 125 Service Manual 2 16...
Страница 32: ......
Страница 62: ......
Страница 64: ......
Страница 80: ......
Страница 82: ...Fluke 123 124 125 Service Manual 5 4 ST8014 EPS Figure 5 1 Fluke 123 124 125 Main Assembly...
Страница 88: ...Fluke 123 124 125 Service Manual 5 10...
Страница 90: ......
Страница 93: ...List of Replaceable Parts 6 4 Final Assembly Parts 6 6 5 ST8014 EPS Figure 6 1 Fluke 123 124 125 Final Assembly...