Performance Verification
3.5 Input A and Input B Tests
3
3-11
•
Fluke 123/124: press
to open the TRIGGER menu, and choose:
INPUT:
B | SCREEN UPDATE:
FREE RUN | AUTO RANGE:
>15Hz
•
Fluke 125: press
to open MENU.
•
Fluke 125: press
to open the TRIGGER menu, and choose :
INPUT:
B | UPDATE:
FREE RUN | AUTO RANGE:
>15Hz
3.
Set the 5500A to source a leveled sine wave of 600 mV peak-to-peak (SCOPE output,
MODE levsine). Set the 5500A to Operate (OPR).
4.
Set the 5500A frequency according to the first test point in Table 3-1.
5.
Observe the Input B main reading on the test tool and check to see if it is within the
range shown under the appropriate column.
6.
Continue through the test points.
7.
When you are finished, set the 5500A to Standby.
3.5.6 Input B Frequency Response Upper Transition Point Test
Proceed as follows to test the Input B frequency response upper transition point:
1.
Connect the test tool to the 5500A as for the previous test (see Figure 3-4).
2.
Select the following test tool setup:
•
Turn Input B on (if not already on).
•
Press
to select auto ranging (
AUTO
in top of display).
Do not press
anymore!
•
Using
change the sensitivity to select manual sensitivity ranging, and lock
the Input B sensitivity on 200 mV/div.
•
Fluke 123/124: press
/
to open the SCOPE INPUTS menu.
•
Fluke 123/124: press
to open the TRIGGER menu, and choose:
INPUT:
B | SCREEN UPDATE:
FREE RUN | AUTO RANGE:
>15Hz
•
Fluke 125: press
to open MENU.
•
Fluke 125: press
to open the TRIGGER menu, and choose :
INPUT:
B | UPDATE:
FREE RUN | AUTO RANGE:
>15Hz
3.
Set the 5500A to source a leveled sine wave of 1.2V peak-to-peak, 50 kHz (SCOPE
output, MODE levsine). Set the 5500Ato Operate (OPR).
4.
Adjust the amplitude of the sine wave to 6 divisions on the test tool display.
5.
Set the 5500A to 20 MHz (Fluke 123) or 40 MHz (Fluke 124/125), without changing the
amplitude.
6.
Observe the Input B trace check to see if it is
≥
4.2 divisions.
7.
When you are finished, set the 5500A to Standby.
Содержание ScopeMeter 123
Страница 6: ......
Страница 12: ......
Страница 16: ......
Страница 28: ...Fluke 123 124 125 Service Manual 2 14 Figure 2 2 Max Input Voltage v s Frequency for VP40 10 1 Voltage Probe...
Страница 30: ...Fluke 123 124 125 Service Manual 2 16...
Страница 32: ......
Страница 62: ......
Страница 64: ......
Страница 80: ......
Страница 82: ...Fluke 123 124 125 Service Manual 5 4 ST8014 EPS Figure 5 1 Fluke 123 124 125 Main Assembly...
Страница 88: ...Fluke 123 124 125 Service Manual 5 10...
Страница 90: ......
Страница 93: ...List of Replaceable Parts 6 4 Final Assembly Parts 6 6 5 ST8014 EPS Figure 6 1 Fluke 123 124 125 Final Assembly...