Fluke RUSKA 2482 Скачать руководство пользователя страница 33

 

Differential Pressure Piston Gauge 

 

Operation 

25 

6.

 

Apply the weights for the next differential point and wait for the Tared Reading to 

stabilize.  

7.

 

When all differential observations are complete, select 

 or 

, remove all the weights, and turn the 

Oil Isolation Valve

 to 

Equalize. Zero differential pressure is now applied to the DUT.  

8.

 

Repeat steps 1 through 7 for each line pressure measurement.  

9.

 

When all measurements are complete, turn the 

Oil Isolation Valve

 to Equalize, 

reduce the line pressure to ambient, select 

 to deactivate the motor 

rotation, and select 

.  

Techniques for Extended Calibrations  

The procedure outlined in Generating Differential Pressures — Hydraulic, relies on the 

Thermal Compensation circuit to control the differential pressure change from one 

differential pressure to the next. During an extended calibration series the capacity of the 

controller may be exceeded. This will be indicated by the inability of the control system 

to establish or maintain the desired differential pressure. When the bar graphs in the 

DeadWeight Controller screen are at their limits (the top bar greater than 80% and the 

bottom bar is less than 20%) and the differential pressure does not come closer to the 

desired level after several minutes, then the controller capacity has likely been exceeded.  

Option 1  

To recover the control capacity press 

, remove all weights, turn the 

Oil 

Isolation Valve

 to Equalize, and wait about 20 minutes before generating further 

differential pressures. It is not necessary to remove the line pressure for control capacity 
recovery.  

Option 2  

The control capacity can also be conserved by using the Hand Pump to bring the 
differential pressure closer to the zero indication. Before changing the masses to obtain a 

new differential pressure, place the controller into 

, then use the hand 

pump to get close to the next Differential pressure. Place the controller back into 

.  

Differential Measurements Using Compressed Gas  

This Section applies when the Test Instrument is tested with compressed gas and is 

connected to the GAS test ports of the 

RUSKA 2482

. Refer to the specifications for the 

requirements for the quality of compressed gas.  

PID selection  

Verify that the Oil/Gas PID option is selected (see the Software section, Figure 4-13, PID 

Values with Control Coefficients).  

Verify Gas Supply Connection  

Verify the proper installation of the gas supply system (see the Installation section).  

Using the Oil/Gas Interfaces  

The oil to gas interface windows allow the maintenance of the oil levels in the system. 

The oil level can be manipulated with the hand pump.  

Содержание RUSKA 2482

Страница 1: ...n USA Specifications are subject to change without notice All product names are trademarks of their respective companies RUSKA 2482 Differential Pressure Piston Gauge Users Manual Via Acquanera 29 221...

Страница 2: ...vice center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB Destination Fluke assumes no...

Страница 3: ...p 9 Software Setup 10 Installation 10 Troubleshooting Communication Setup 11 WinPrompt Software 11 RUSKA 2482 Software 12 Menu Commands 13 Menu File Exit 13 Menu Communications Port 14 Menu Communicat...

Страница 4: ...nded Calibrations 28 Head Height 29 Maintenance 30 Pump 30 Screw Press Seals 32 Complete Disassembly 33 Full System Purge of Air 33 Major Purge Process 34 Minor Purge Process 34 Balance Removal Instal...

Страница 5: ...iii List of Tables Table Title Page 1 Symbols 2 2 Oil and Gas Isolation Valves States 26 3 Reservoir Test Port Assemblies and Screw Press Assembly Parts List 31...

Страница 6: ...RUSKA 2482 Users Manual iv...

Страница 7: ...troller Current Status Window 12 13 File Menu 13 14 Communications Menu 14 15 Comm Port Selection 14 16 Display Resolution Selection 15 17 Line Pressure Units Selection 15 18 Balance Filter Selection...

Страница 8: ...RUSKA 2482 Users Manual vi...

Страница 9: ...general safety precautions that are not related to any specific procedures and do not appear elsewhere in this publication These are recommended precautions that personnel must understand and apply d...

Страница 10: ...ebsite for recycling information P This equipment meets the requirements of all relevant European safety directives The equipment carries the CE mark General Information The RUSKA 2482 Differential Pr...

Страница 11: ...n Program NVLAP for conformance with criteria set forth in NIST Handbook 150 2001 and all requirements of ISO IEC 17025 1999 A NVLAP accredited calibration certificate is provided standard with the RU...

Страница 12: ...open Changes in the line pressure may result in a small offset on the Balance This effect is nulled by using the control program to tare the offset and establish a zero datum at the given line pressur...

Страница 13: ...sturdy workbench able to safely support the additional weight of the RUSKA 2482 with accessories approximately 45 kg 100 lbs PC computer with Microsoft Windows 2000 or XP 400 MHz 1 6 MB hard disk spac...

Страница 14: ...the left panel door two from the rear panel and one from the bottom side panel b Open the panel door c Loosen and remove the balance transit strap d Close the side panel door e Replace the three pane...

Страница 15: ...KA 2482 rear panel and the tee fitting located on the rear of the hydraulic pump assembly see Figure 5 7 Route the waste oil drain tube to a suitable container compatible with Dioctyl Sebacate Oil DOS...

Страница 16: ...sition While the Oil Isolation Valve is in Equalize mode turning the pump spindle generates line pressures While the Oil Isolation Valve is in Isolation mode turning the pump spindle will generate Del...

Страница 17: ...anel of the instrument see Figure 5 gms09 bmp Figure 8 Power Connection 2 Plug the power adapter cord into a suitable wall outlet The power lamp on the front panel of the RUSKA 2482 will turn on 3 Con...

Страница 18: ...inPrompt CD Version 3R00 or later into the computer If Autorun does not begin the installation then manually activate the setup exe program on the CD Refer to the WinPrompt manual for more details Sin...

Страница 19: ...ght through cable 3 Verify that the power is applied to the RUSKA 2482 instrument The LED on the RUSKA 2482 instrument will turn on when power is applied WinPrompt Software WinPrompt performs pressure...

Страница 20: ...current state and provides buttons for the major operations of the RUSKA 2482 It also provides for setup calibration and opening of all other windows available in the RUSKA 2482 The Current Status wi...

Страница 21: ...be allowed to stabilize in this mode before attempting Control mode Control Temperature controllers are actively controlling in order to bring the Tared Reading to Setpoint Nudge Setpoint Adjusts the...

Страница 22: ...enu Click the drop down arrow next to the current selection to see the available communication ports If the port is not listed it is not installed correctly in Microsoft Windows or is not connected to...

Страница 23: ...the units for the Line Pressure displayed in the Main window gms18 bmp Figure 17 Line Pressure Units Selection Menu Set up Balance Filter Select the Balance Filter to turn on and off If a check mark...

Страница 24: ...A 2482 Users Manual 16 gms19 bmp Figure 18 Balance Filter Selection Menu Set up Calibrate Line Displays the dialog for calibration of the line pressure transducer gms20 bmp Figure 19 Calibration Selec...

Страница 25: ...the Actual pressure 4 Click The C0 coefficient will be adjusted to give the correct pressure reading 5 Click To calibrate the line sensor Set up Calibrate Line 1 Click 2 Apply the first known pressure...

Страница 26: ...ave The Coefficient File options allow the user to Save and Load all of the calibration and setup of the RUSKA 2482 software Selecting Save will write the coefficients to a user selected filename Sele...

Страница 27: ...instrument controller There are two sets of Fine coefficients one for Oil calibrations and a second for Oil Gas calibrations The Coarse coefficients are used for both calibration types These coeffici...

Страница 28: ...onal filtering of the Balance Reading Note Setup Balance Filter must be checked for the filter to be used The filter time constant has the units of seconds and can be entered in 0 1 increments Most sy...

Страница 29: ...communication problems When the instrument is operating normally this window can be closed gms27 bmp Figure 26 RS 232 View Development Displays in a hexadecimal format the serial communication strings...

Страница 30: ...lear Tare and with the motor turned on then the Tared Reading will be equivalent to the total mass of the piston and mass table 935 grams Since the Balance is used as a reference device the 935 grams...

Страница 31: ...g used to connect the Device Under Test to the RUSKA 2482 must be rated for safe operation to at least 2975 psi 205 bar Purge the air from the interconnect tubing Put the Oil Isolation Valve in Equali...

Страница 32: ...e Oil Isolation Valve to Isolate apply weights corresponding to the desired differential pressure and press When the control system regulates the differential pressure for a Tared Reading of zero then...

Страница 33: ...closer to the desired level after several minutes then the controller capacity has likely been exceeded Option 1 To recover the control capacity press remove all weights turn the Oil Isolation Valve...

Страница 34: ...be in the four possible states listed in the table below Table 2 Oil and Gas Isolation Valves States State Oil Isolation Valve Gas Isolation Valve Equalize Zero P Equalize Equalize Caution Avoid Equal...

Страница 35: ...section Windows PIDValue Operation Power This will take approximately 15 minutes prior to a calibration sequence to provide the maximum control capacity for the system This may be unnecessary for sho...

Страница 36: ...nts are complete reduce the line pressure to ambient by carefully opening the Gas Exhaust Metering Valve Turn off the motor by pressing Techniques for Extended Calibrations During an extended calibrat...

Страница 37: ...l when the zero is performed but the oil levels remain constant during the calibration series then the head height does not matter as it is incorporated in the zero setting The only hydraulic head pre...

Страница 38: ...levant part numbers Where ASSY appears as a part number this component is associated with other components in an assembly for replacement purposes Before beginning any maintenance remove any instrumen...

Страница 39: ...Test Port Assemblies and Screw Press Assembly Parts List ITEM DESCRIPTION PART 1 Banjo Bolt P101230 2 O Ring P103614 3 Banjo Test Port P101226 4 O Ring P103606 5 O Ring 54 703 12 6 Valve Body SDW5790...

Страница 40: ...The white anti extrusion ring 21 is a PTFE spiral and can be removed by unwinding it from the rambler 5 Do not use a sharp tool that will scratch the surfaces of the rambler when removing the rambler...

Страница 41: ...ing barrel Screw the barrel fully into the test station and secure with the locknut 10 Inspect all seals for wear or damage Replace as necessary Note If the rambler shows signs of wear the main bore w...

Страница 42: ...weight table and hold it at the bottom of travel W Caution Do not allow the weight table to return to a normal position while the low port shut off valve is open 10 Close the low port shut off valve 1...

Страница 43: ...d put the Oil Isolation Valve to Isolate Mode This will keep the Piston Spring clear of the Balance 6 Remove the screws from the front and rear of the Balance mounting plate 7 Slide the Balance and Mo...

Страница 44: ...conditions Differential Pressure Reproducibility Differential Pressure Reproducibility 4 5 Pa1 Refer to the Calibration Report for Pressure Uncertainty Piston and Cylinder Thermal Coefficient piston...

Страница 45: ...of oil 3 Verify that line pressure is removed 4 Verify that all ports are sealed to minimize oil loss Installation of Clamps Install the Balance Strap and Transit Clamp as shown in the Installation s...

Страница 46: ...RUSKA 2482 instrument 2482 107 Shipping case for RUSKA 2482 accessory Final Instructions Include in the shipment Statement of the problem or service required The name and contact information of a know...

Отзывы: