Differential Pressure Piston Gauge
Software Setup
13
The Compensator control loop is turned on and off by these buttons:
Button Action
Off
No compensation, no heating nor cooling is done.
Standby
The temperature controllers are being biased to defined presets. The
heating is constant and not affected by the Tared Reading. This mode
provides increased control distance in one direction by biasing. Temperature
should be allowed to stabilize in this mode before attempting Control mode.
Control
Temperature controllers
are actively controlling in order to bring the Tared
Reading to Setpoint
.
Nudge Setpoint
Adjusts the setpoint in small increments
Motor
Turns motor on (green) and off (red).
Tare
Zeros the Tared Reading by changing the tare value. This is used to remove
offsets in the Tared Reading while the piston is equalized.
Menu Commands
Menu: File | Exit
When running as a standalone program, File
→
Exit closes the application. A warning will appear
if attempting to exit while the controller is running.
gms14.bmp
Figure 13. File Menu