2482
Tared Reading is
transferred real time to the
active WinPrompt Calibration
Point. This Trim is used real
time to adjust the value of
the Actual Pressure.
gms12.eps
Figure 11. RUSKA 2482 - WinPrompt® - Typical Setup
A typical setup for the RUSKA 2482 is shown in Figure 11. Above left is the
WinPrompt
®
window; above right is the chart window, and below right is the
RUSKA 2482 deadweight controller window
RUSKA 2482 Software
The RUSKA 2482 Deadweight Controller window is the primary operation window of
the RUSKA 2482 software. It shows the current state and provides buttons for the major
operations of the RUSKA 2482. It also provides for setup, calibration, and opening of all
other windows available in the RUSKA 2482.
The Current Status window (see Figure 12) shows the following items:
Current tared reading from the internal balance.
This reading can be zeroed by clicking.
This value is transferred to WinPrompt and is
used as the Trim Value for thecurrent pressure
calculation.
Line pressure sensor reading.
Current Tare value. This value is set by clicking
to make the tared reading zero.
Setpoint is the targeted grams value for the
thermal controller.
Shows the power setting being applied to each
of the Temperature Controllers.
gms13.eps
Figure 12. RUSKA 2482 Deadweight Controller - Current Status Window