Differential Pressure Piston Gauge
NVLAP Accredited Calibrations
3
gms02.bmp
Figure 1. Basic System Diagram
NVLAP Accredited Calibrations
Fluke is recognized by the United States Department of Commerce National Institute of
Standards and Technology’s National Voluntary Accreditation Program (NVLAP) for
conformance with criteria set forth in NIST Handbook 150:2001 and all requirements of
ISO/IEC 17025:1999.
A NVLAP accredited calibration certificate is provided standard with the RUSKA 2482.
Theory of Operation
An overview of the RUSKA 2482 system is shown in Figure 2. During the initial
configuration of the system, the Balance is internally aligned and zeroed. With only the
weight of the piston and associated hardware acting on the Balance, the Balance is zeroed.