Table 3. T+ and T+ PRO Accuracy Tests (cont.)
Step Function
Range
Input
Value
UUT Response
(LED)
LCD Display (T+ PRO)
Measurement Limits
14
Voltage LEDs
24 V
24 V
24 V LED On
- - -
15
120 V
120 V
120 V LED On
- - -
16
208 V
208 V
208 V LED On
- - -
17
240 V
240 V
240 V LED On
- - -
18
277 V
277 V
277 V LED On
- - -
19
347 V
(Canada)
347 V
347 V LED On
- - -
20
480 V
480 V
480 V LED On
- - -
21
Continuity
0 – 20 k
Ω
Short
Continuity LED On
Continuity Sign
22
Resistance
0.00 – 9.99 k
Ω
Short
None
0.00 to 0.03
23 100
Ω
None
0.06 to 0.14
24 9
k
Ω
None
8.52 to 9.48
Load Current Test
To test the load current, set up the UUT with equipment as shown in Figure 2 and complete the
following procedure:
1.
Set the DMM to the mA dc function.
2.
Apply 600 V dc from the Source. The DMM reading should be < 5 mA.
3.
Apply
−
600 V dc from the Source. The DMM reading should be < 5 mA.