![Fluke E1M Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/fluke/e1m/e1m_user-manual_2309862051.webp)
Operation
Menu Structure
51
6
TOPSENSOR OFFSET
The subentry is only available for sensors in 2-color mode. It allows to correct the wide band temperature reading
by the addition of an offset value in a range from -200.0 to 200.0°C. The standard value is 0.0°C.
MATCH
adapts the temperature reading to the expected real object temperature. You can affect the current temperature
reading by overriding it with the real, alternatively measured, object temperature. In 1-color mode, the match
confirmation corrects the object emissivity value to match the current temperature reading. The match confirmation
in 2-color
mode adapts the slope value to match the current temperature reading. With the ▲ ▼ keys, you can
toggle between temperature match values from “LOW LIMIT” to “HIGH LIMIT” of the temperature range.
AVERAGE
defines the parameter for the signal post processing, see section 6.3.1
PEAK HOLD
defines the parameter for the signal post processing, see section 6.3.2
VALUE HOLD
defines the parameter for the signal post processing, see section 6.3.4
DECAY RATE
defines the linear signal decay for a given time span. The unit for the decay is in °C/s or °F/s, see
in section 6.3.2.3
ALARM SOURCE
sets the source for an alarm:
•
T
obj
– the object temperature is the alarm source
•
T
int
– the internal temperature of the sensor is the alarm source
SETPOINT
Once the setpoint is activated by providing a non-zero value, the relay changes state as the current temperature
passes the setpoint temperature. With the ▲ ▼ keys, you can toggle between setpoint values from “LOW LIMIT”
to “HIGH LIMIT” (e.g. 400.0 °C to 1800.0 °C).
For more information, see section 5.7.1.1
DEADBAND
defines the deadband, see section 5.7.1.2
BACKGROUND CONTR
refers to the compensation source for the background temperature, see section 5.6.2.2
•
NO COMPENSATION
– background temperature compensation is switched off
•
EXTERNAL mA IN
– temperature value for background compensation is taken from the external current
loop analog input
•
TEMP. VALUE
– temperature value for background compensation is taken from subentry BACKGROUND
TEMP.
BACKGROUND TEMP.
provides the temperature value for the background compensation.
With the ▲ ▼ keys, you can toggle between
values between
“LOW LIMIT” to “HIGH LIMIT” of the temperature measurement range.
Содержание E1M
Страница 113: ...Accessories Mechanical Accessories 113 12...