271
Users Manual
7-2
Introduction
In burst mode a defined number of cycles follow each trigger event. This mode is edge
triggered.
In gated mode the generator runs while the gating signal is high. This mode is level
sensitive.
Both burst and gated modes can be controlled by either the internal trigger generator, an
external trigger input, by the front panel
MAN/SYNC
key or by remote control.
Internal Trigger Generator
The internal trigger generator divides down a crystal oscillator to produce a 1:1 square-
wave with a period from 0.02 ms (50 kHz) to 200 s (0.005 Hz). Generator period entries
that cannot be exactly set are accepted and rounded up to the nearest available value, e.g.
0.109 ms is rounded to 0.12 ms. The generator output is available as a TTL level signal at
the rear panel
TRIG/SWEEP OUT
socket.
In burst most the rising edge of each cycle of the trigger generator is used to initiate a
burst; the interval between bursts is therefore 0.02 ms to 200 s as set by the generator
period.
In gated mode the output of the main generator is gated on whilst the trigger generator
output is high; the duration of the gate is therefore 0.01 ms to 100 s in step with trigger
generator periods of 0.02 ms to 200 s.
External Trigger Input
External trigger or gate signals are applied to the front panel
EXT TRIG
input which has
a TTL level (1.5 V) threshold. In triggered burst mode the input is edge sensitive; the
rising edge of each external trigger initiates the specified burst. In gated mode the input is
level sensitive; the output of the main generator is on whilst the gate signal is high
(>1.5 V).
The minimum pulse width that can be used with the
EXT TRIG
input is 50 ns and the
maximum repetition rate is 1 MHz. The maximum signal level that can be applied
without damage is ±10 V.
Triggered Burst
Triggered burst mode is turned on and off with alternate presses of the
TRIG
key; the
lamp beside the key lights when triggered mode is on. The triggered mode parameters
(trigger source, internal trigger generator, burst count and start/stop phase) are all set
from the trigger edit menu which is selected by pressing the
EDIT
key followed by the
TRIG
key. When trigger edit is selected the lamp beside the
TRIG
key flashes to show
edit mode regardless of whether triggered burst operation is currently selected to be on or
off.
SOURCE=EXT
TGEN=1.00ms 1.000kHz
BURST COUNT= 0001
PHASE=+000° (+000°)
1.888.475.5235
Fluke-Direct
.com
Содержание DDS Function Generator 271
Страница 12: ...271 Users Manual x 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 22: ...271 Users Manual 1 10 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 34: ...271 Users Manual 4 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 46: ...271 Users Manual 5 12 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 52: ...271 Users Manual 6 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 60: ...271 Users Manual 8 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 64: ...271 Users Manual 9 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 76: ...271 Users Manual 12 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 80: ...271 Users Manual 13 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 90: ...271 Users Manual 15 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 110: ...271 Users Manual 16 20 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 112: ...271 Users Manual 2 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 127: ...Appendix G Front and Rear Panels Front panel sha0025f gif G 1 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 128: ...271 Users Manual Rear panel sha0027f gif G 2 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 132: ...4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...