271
Users Manual
1-8
Inputs
Ext Trig
Frequency range:
dc to 1 MHz.
Signal range:
Threshold nominally TTL level; maximum input ±10 V.
Minimum pulse width:
50 ns for trigger, gate and FSK modes;
1 ms for sweep and hop modes.
Input impedance:
10 k
Ω
VCA In
Frequency range:
DC - 100 kHz.
Signal range:
2
.
5 V for 100% level change at maximum output.
Input impedance:
typically 6 k
Ω
.
Phase locking
The signals from these sockets are used to phase lock two or more generators.
Clock In/Out
TTL/CMOS threshold level as an input.
Output logic levels nominally 1 V and 4 V from typically 50
Ω
as an output.
Sync Out
TTL/CMOS logic levels from typically 50
Ω
.
Interfaces
Full remote control facilities are available through the RS232 or GPIB interfaces.
RS232:
Variable Baud rate, 9600 Baud maximum. 9-pin D-
connector.
As well as operating in a conventional RS232 mode the
interface can be operated in addressable mode whereby up to
32 instruments can be addressed from one RS232 port.
GPIB (IEEE-488):
Conforms with IEEE488.1 and IEEE488.2
General
Display:
20 character x 4 row alphanumeric LCD.
Data Entry:
Keyboard selection of mode, waveform etc.
Value entry direct by numeric keys or by rotary control.
Stored Settings:
Up to 9 complete instrument set-ups may be stored and
recalled from battery-backed memory.
Size: 3U
(130mm)
high;
half-rack (212mm) wide;
330mm deep.
Weight:
4.1 kg (9 lb)
Power:
100 V ac, 110-120 V ac or 220-240 V ac ±10 %,
50/60 Hz, adjustable internally;
40 VA max.
Installation category II.
Temperature range:
operating: +5 to 40 °C, 20-80 % RH.
storage: -20 to +60 °C
Environmental:
Indoor use at altitudes up to 2000 m,
1.888.475.5235
Fluke-Direct
.com
Содержание DDS Function Generator 271
Страница 12: ...271 Users Manual x 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 22: ...271 Users Manual 1 10 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 34: ...271 Users Manual 4 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 46: ...271 Users Manual 5 12 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 52: ...271 Users Manual 6 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 60: ...271 Users Manual 8 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 64: ...271 Users Manual 9 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 76: ...271 Users Manual 12 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 80: ...271 Users Manual 13 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 90: ...271 Users Manual 15 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 110: ...271 Users Manual 16 20 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 112: ...271 Users Manual 2 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 127: ...Appendix G Front and Rear Panels Front panel sha0025f gif G 1 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 128: ...271 Users Manual Rear panel sha0027f gif G 2 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 132: ...4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...