1
Temperature and storage range with Energizer Lithium L91 cells is expanded to -30 to +60°C (-22°F to
+140°F)
2
To connect to two external transducers, a Multi-Parameter 2-Way Adaptor Box (RF615B) is required. Up to 4
external channels can be achieved by connecting a Y-Adaptor (RF602Y) with two Multi-Parameter 2-Way
Adaptor Boxes (RF615B).
External Probes Specifications
Number of External Channels
4
Sensor Type
Voltage or Current Measurement
Connector
6-Pin Lumberg
2
Scales
Free Form on Cloud
Measurement Range
+/-0.32000 to +/-32000
Decimal Points
1DP to 5DP
Displayed Range (LCD Only) Not Applicable to logged readings
+/-0.32 to +/-19999
Instrument Accuracy
+/-0.3% of reading @ +23°C Ambient
Resolution
Resolution @ 0-10V +/-1mV
Resolution @ 0-1V +/-0.1mV
Resolution @ 4-20mA 1µA