The Diligence 600 WiFi Monitoring System will alarm in several different ways and,
depending on the licence applied to the
Diligence Cloud
platform, there are two types of
Temperature Alarm
for Diligence 600 transmitters.
High and Low Alarms
are traditional alarms which will generate the alerts/emails from the
Diligence Cloud
. A maximum and minimum alarm level should be set to ensure that alarm
events can be triggered.
High and Low Alerts
are levels that sit below and above the absolute
High and Low Alarm
levels. When these
High and Low Alerts
are triggered, a warning is added to indicate that
your monitored area is heading towards an alarm excursion. At this stage it is not yet in
alarm.
High and Low Alerts
are available with
Diligence Cloud
licence plans above
Lite
level.
They remain optional and do not have to be set.
An
Alarm Delay
can be set for both
High and Low Alarms
as well as
High and Low Alerts
.
This
Alarm Delay
can be a value from 0 to 60 minutes and provides a simple delay to the
alarm/alert being triggered. This allows for small deviations, above or below an alarm or
alert value, to be ignored until they have been in evidence for the given period at which point
the alarm/alert is then required. The
Alarm Delay
is an optional setting and does not have to
be programmed to the transmitter.
Temperature Alarm Limits
are set by the maximum and minimum measurable limits of the
transmitter. The
Task
for the transmitter will not allow you to set a limit outside of the
measurement range for the channel. Please refer to the specifications for your transmitter.
22
Copyright © 2022 Comark Instruments