BT521
Users Manual
4-2
Table 4-1. Elements of the Interactive Test Probe
Item No.
Name
Function
Display
Shows information such as measurement readings and battery
ID.
LED
indicator
Indicates the status of a measurement. Green means Pass;
orange means Warning; red means Fail.
Save button
Manually saves a measurement reading.
Backlight key
Turns on or turns off the backlight for the interactive test probe.
Power key
Turns on or turns off the interactive test probe.
Speaker
Plays audio reminders (PASS, WARNING, FAIL, and battery
number) when enabled.
Probe
tip
Connects the interactive test probe to the measured unit. The
probe tips can be replaced.
Tactile
barrier
Hold the handle behind the tactile barrier to prevent personal
injury.
Connect the Probe to the Product
Before the interactive test probe is used, connect the probe to the Product.
Configure the Interactive Test Probe on the Product
Set the Audio
The interactive test probe has a speaker that can play audio reminders.
To set the audio:
1. Push
to open the Setup screen.
2. Use
L
to highlight
Handle
, and push the
Select
softkey.
Audio
is already highlighted.
3. Push the
Select
softkey.
4. Use
and
L
to highlight
Disable
,
Threshold result
,
Battery number
, or
Both
.
5. Push the
Confirm
softkey.
6. Push the
Back
softkey to return to the Setup screen.
Set Temperature Unit
To set the temperature unit for the interactive test probe:
1. Push
to open the Setup screen.
2. Use
L
to highlight
Handle
, and push the
Select
softkey.
3. Use
L
to highlight
Temperature unit
, and push the
Select
softkey.
4. Use
and
L
to highlight
Celsius
or
Fahrenheit
, and push the
Confirm
softkey.
5. Push the
Back
softkey to return to the Setup screen.