Use the Product
Make Measurements
3
3-17
Measure Current (AUX Function)
The Product can measure current (dc and ac) contactless with the BTL_A adapter and
the Fluke i410 Current Clamp.
To measure current, turn the rotary switch to
AUX
. See Figure 3-5.
hsz054.eps
Figure 3-5. AUX function
Set Measurement Range
In AUX function, both ac current and dc current use a fixed 400 A range. This cannot be
changed.
The frequency for ac current uses auto range. This cannot be changed either.
Save Current Measurement Readings
In Meter mode, push the
Save
softkey to save the current reading (including frequency
in ac current) and the timestamp. All saved data is stored in chorological order.
In Sequence mode, push the
Save
softkey to save the current reading (including
frequency in ac current) and the timestamp. The current battery number automatically
increases by 1. The cell that corresponds to the tested battery becomes full, and the
cursors move forward.
Note
Up to 20 current readings can be saved in one profile.
Use the i410 Current Clamp
For details about how to use the i410 Current Clamp, refer to the separate
i410
Instruction Sheet
supplied with the Product.
1.888.610.7664
Fluke-Direct
.com