Use the Product
Make Measurements
3
3-11
The Product lets you define upper and lower measurement thresholds or tolerance
ranges. These defined threshold values are then compared to the measured values to
automatically identify and prompt the user with a
PASS
,
FAIL
or
WARN
indicator of
battery out of tolerance conditions.
The threshold function is disabled by default. You can set up to 10 set of thresholds and
select one threshold as needed.
To set and select measurement thresholds:
1.
On the measurement screen, push the
More
softkey and
Threshold
softkey to
open the Select Threshold menu.
2. Use
and
to select one threshold set out of ten.
hsz033.jpg
3. Use
and L to highlight the value to be edited among
Voltage lower
,
Reference
,
Warning
,
Fail
,
Low limit
, and
Notes
.
4.
Edit the selected field.
a. Use
the
–
and
+
softkeys to change the values for
Warning
and
Fail
.
b.
For other fields, push the
Edit
softkey, use the arrow keys to edit the value,
and then push the
Confirm
softkey to save the value.
5.
Once all threshold values are correct, push the
Confirm
softkey on the save the
threshold set.
The threshold set is applied and the
T-X
(X stands for value of
No
.) icon and the
corresponding PASS/WARN/FAIL indication shows on the display.
To disable measurement thresholds:
1.
On the measurement screen, push the
Threshold
softkey to open the Select
Threshold menu.
The value of
No.
is already highlighted.
2. Use
to set
No.
to
---
.
3. Push
the
Confirm
softkey.
The
T-X
icon no longer shows on the display.