7526A
Calibration Manual
18
Ohms Output
The ohms output performance test verifies the accuracy of the ohms output from the front
of the Product. Use the DMM to measure the ohms output. Apply the values shown in
Table 7. See Figure 6 for connections.
Note
Make sure the sense wires from the 8508A are on the bottom of the
stackable leads.
Push
to select the 400
Ω
, then the 4000
Ω
range.
Note
The 8508A should be set to the 2 k
Ω
range when you measure the 7526A
400
Ω
range. After the 1000
Ω
measurement, the 8508A should be set to the
20 k
Ω
range with loI off. To prevent an overload, the 8508A must be in the
“loI” (low current mode) to measure 5
Ω
in the 4 k
Ω
range.
CURRENT
OUTPUT
HI
VOLTS
mA
RTD/
OUTPUT
HI
LO
HI
LO
LO
TC
SENSE
INPUT/OUTPUT
4W RTD/
INPUT
20 V PK
MAX
100 V MAX
100 mA MAX
20 V PK
MAX
20 V PK
MAX
PRECISION PROCESS CALIBRATOR
7526A
HI
8508A
7526A
gxm014.eps
Figure 6. Resistance Output Performance Test Connections
Table 7. Ohms Output Ranges
Range
Output
90-Day Specification
1-Year Specification
400
Ω
5
Ω
0.012
Ω
0.015
Ω
100
Ω
0.012
Ω
0.015
Ω
200
Ω
0.012
Ω
0.015
Ω
300
Ω
0.012
Ω
0.015
Ω
400
Ω
0.012
Ω
0.015
Ω
4000
Ω
5
Ω
0.25
Ω
0.30
Ω
1000
Ω
0.25
Ω
0.30
Ω
2000
Ω
0.25
Ω
0.30
Ω
3000
Ω
0.25
Ω
0.30
Ω
4000
Ω
0.25
Ω
0.30
Ω