Automatic Pressure Calibrator
HART Functionality
25
HART Functionality
The Product interfaces with HART devices. It can:
•
View and modify transmitter setup and data that includes:
◦
Write Lower Range Value (LRV)
◦
Write Upper Range Value (URV)
◦
Write PV Unit
◦
Write Message and Descriptor
◦
Write Damping Value
◦
Write Date
•
Perform service features including:
◦
PV Zero Trim
◦
mA Output Trim
◦
Trim to Applied Values
◦
Set Fixed mA Output
◦
Re-range Transmitter
◦
Device Diagnostics
•
Perform HART calibration without using a Task (Ad hoc)
•
Perform HART calibration from a downloaded Task
Push
G
and the Product searches (polls) the bus to find any connected HART
devices. From this screen, there are these softkeys:
A
-
Communicator Only
B
-
Loop Power +24V ON
C
-
Enable HART 250
Ω
D
-
Restart BUS Polling
Once Polling finds the transmitters, move the cursor to the tag number and push
D
(
Continue
) to select.