![Fluke 729 Pro Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/fluke/729-pro/729-pro_user-manual_2310413029.webp)
Automatic Pressure Calibrator
Tasks Menu
23
Pressure Transmitter (P/V)
Use this feature to calibrate a pressure to voltage transmitter with voltage measurement.
For connections, see Figure
1. From page 1 of the Tasks menu, push the arrows to highlight
Pressure Transmitter
(P/V)
.
2. Push
E
or
R
.
3. From this screen, use the arrows, numeric keypad, and
E
to change the values of
the source pressure and measure voltage.
4. Push
D
(
Continue
) to edit these other parameters:
•
Tolerance of Range
•
Test Strategy
•
Auto Settling Time
•
Transfer Function
A
(
Linear or Square Root
).
5. Push
B
(
Save as Customized
) to save these settings as a customized task.
Recall these customized tasks from the Tasks menu at any time.