6100B/6105A
Getting Started
24
Rear-Panel Features
Figure 6 and Table 4 show and explain the rear panel features.
A
B
C
D
E
F
G
H
I
J
K
L
M
N
O
P
gdw07.eps
Figure 14. Rear-Panel Features
Table 4. Rear-Panel Features
Number Meaning
Main Power On-Off Switch -This is a true mains isolating switch.
Auxiliary Unit Control Connectors - Connection to Auxiliary units via
Fluke supplied cable.
Trigger Out Connector - The Trigger Output Connector has a +5 V
CMOS logic drive providing a falling edge time marker intended to
synchronize external equipment to the dip/swell function. The point at
which the falling edge occurs is controlled by the Trigger Output Delay.
After the falling edge the signal will remain low for a minimum of 10
μ
s.
Trigger Input Connector -iA TTL compatible input which can be
selected to initiate a dip/swell on a falling edge. The falling edge can
either start the user programmable initial delay timer or arms the user
settable output waveform phase angle comparator. These are mutually
exclusive. When the timer delay has expired or the comparator has
found the required angle of the output waveform the Ramp In section of
the dip/swell will commence. The input must remain low for 10
μ
s after
the falling edge to be recognized properly.