Electrical Power Standards
6100B Set Up Exercise
19
1.
Push the
Enable/Disable
softkey and the L1 current row in the Output Menu shows
as “Enabled”. See Figure 7.
gdw09.bmp
Figure 7. The Output Menu (L1 Current Selected)
The Current Channel is Enabled but not Operating
There is not yet output at the 6100B output terminals. The system must be set to operate
mode to achieve this. If the system was set to operate at this point, nothing would appear
at the voltage terminals because that channel is not enabled. The system would attempt to
output 0.5 A from the current terminals but this would generate a warning message as the
output is open-circuit. This will be addressed once the Current channel is set up.
Set Up the L1 Current Channel
To change the settings for an output channel first move the Menu focus to the Waveform
menu.
1.
Push
twice. The Menu focus moves to the Waveform menu. See Figure 8.
gdw11.bmp
Figure 8. Waveform Menu, Fundamental
2.
The Range control has focus (blue background).
3.
Use the
and
or the scroll wheel to set the range to 2 A. Push
or
to
complete the entry and move focus to the RMS field.
4.
Set the output RMS value to 2.0 A using the controls as described in “Data Entry