568 EX
Users Manual
20
Specifications
General Specifications
IR Temperature
Range
-40
°
C to 800
°
C (-40
°
F to 1472
°
F)
Accuracy
<0 °C: ±(1.0 °C + 0.1 °/1 °C)
≥
0 °C: ±1 % or ± 1.0 °C, whichever is greater
<32 °F: ±2 °F ±0.1 °/1 °F
≥
32 °F: ±1 % or ±2 °F, whichever is greater
Repeatability
±
0.5 % of reading or
±
0.5 °C (1 °F), whichever is
greater.
Display Resolution
0.1
°
C / 0.1
°
F
Spectral Response
8
µ
m to 14
µ
m
Response Time
<500 ms (95 %)
KTC Input Range
-270
°
C to 1372
°
C (-454
°
F to 2501
°
F)
KTC Input
Accuracy
<-40
°
C:
±
(1 °C + 0.2 °/1 °C)
≥
-40
°
C:
±
1 % or 1 °C, whichever is greater
<-40
°
F:
±
(2 °F + 0.2 °/1 °F)
≥
-40
°
F:
±
1 % or 2 °F, whichever is greater
KTC Resolution
0.1
°
C/0.1
°
F
Distance: Spot
50:1 (90 % energy)
Laser sighting
Single laser, output <1 mW Class II,
wavelength 630 to 670 nm
Emissivity
Digitally adjustable from 0.10 to 1.00 by 0.01 or via
built-in table of common materials
Data storage
99 points
Operating Altitude
2000 meters above mean sea level
Storage Altitude
12,000 meters above mean sea level
Relative Humidity
10 % to 90 % RH non-condensing up to 30 °C
(86 °F)
Operating
Temperature
0 °C to 50 °C (32 °F to 122 °F)