42
Command List
(cont)
UUT_SET?
Responses:
1. (Integer) 300, 600, 1200, 2400,
4800, or 9600 (baud rate)
2. (Character) XON, RTS, or
NOSTALL (stall method)
4. (Character) DBIT7 or DBIT8
(number of data bits)
5. (Character) SBIT1 or SBIT2
(number of stop bits)
6. (Character) PNONE, PEVEN, or
PODD (parity)
VAL?
Parameter:
(Optional) Units to return
Response:
1. (Float) Measured temperature or
pressure
2. (Character) CEL, FAR, OVER,
OPENTC, OHM, F, PSI, MHG,
INHG, INH2O, FTH2O, MH2O,
BAR, PAL, G/CM2, INH2O60F, or
NONE
VVAL?
Response: (Float)
TC measurement as a
voltage
*WAI
Parameters: None
WAVE
Parameters: 1. SINE, TRI, SQUARE, TRUNCS,
or NONE (Waveform)
2. (Optional) SINE, TRI, SQUARE,
or TRUNCS (Second waveform
for dual outputs)