Command List
(cont)
UNCERT?
Parameter:
1. (Optional) Preferred unit of
primary output uncertainty or PCT
(default)
2. (Optional) Preferred unit of
secondary output uncertainty or
PCT (default)
Response:
1. (Float) 90-day specified
uncertainty of primary unit
2. (Float) 1-year specified
uncertainty of primary output
3. (Character) Unit of primary output
uncertainty
4. (Float) 90-day specified
uncertainty of secondary unit
5. (Float) 1-year specified
uncertainty of secondary output
6. (Character) Unit of secondary
output uncertainty
UUT_FLUSH
Parameter:
None
UUT_RECV?
Response:
(String) Data from the UUT serial
port
UUT_RECVB?
Parameter:
(Optional) Maximum number of
integers per line
Response:
(Indefinite ASCII) Comma separated
integers as follows:
1. (Integer) Number of data bytes
returned excluding the count
2. (Integer) Data from the UUT serial
port as series of comma
separated integers