27
Command List
(cont)
DBMZ_D?
Response:
(Character) Z50, Z75, Z90, Z100,
Z135, Z150, Z300, Z600, Z900,
Z1000, or Z1200
DC_OFFSET
Parameters: Offset amplitude with optional
multiplier and voltage unit
DC_OFFSET?
Response:
(Float) Signed offset amplitude
DPF
Parameters: 1. Cosine of phase offset (0.0 to 1.0)
2. (Optional) LEAD (default) or LAG
DPF?
Responses:
1. (Float) Cosine of phase
offset
2. (Character) LEAD or LAG
DUTY
Parameter:
Duty cycle percentage (.1 to 99.9)
DUTY?
Response:
(Float) Duty cycle
EARTH
Parameter:
OPEN or TIED
EARTH?
Response:
(Character) OPEN or TIED
EDIT
Parameter:
PRI, SEC. FREQ, or OFF
EDIT?
Response:
(Character) PRI, SEC. FREQ, or
OFF
ERR?
Responses:
1. (Integer) Error code
2. (String) Error message string,
formatted for display
ERR_UNIT
Parameter:
GT1000, GT100, GT10, PPM, or
PCT