33
Command List
(cont)
OUT?
Parameters: (Optional) Units of returned
amplitudes:
ac voltage mode (V or DBM)
RTD mode (CEL, FAR, or OHM)
TC output mode (CEL, FAR, or V)
Responses:
1. (Float) Output amplitude
2. (Character) Units (V, DBM, A,
OHM, F, CEL, or FAR)
3. (Float) Second output amplitude
for dual output functions (0 if no
second amplitude)
4. (Character) Units of second
amplitude (0 if no second
amplitude)
5. (Float) Frequency (0 if dc,
resistance, capacitance, or
temperature)
OUT_ERR?
Responses:
1. (Float) UUT error magnitude
2. (Character) Units for the above
number (PPM, PCT, DB, or 0)
PHASE
Parameter:
Phase with optional multiplier
and
DEG unit
PHASE?
Response:
(Float) Phase in degrees
POWER?
Response:
(Float) Power in ac or dc power