5320A Users 

Manual Supplement 

4/12 3 

Change #5 

On page 6-12, replace 

Table 6-8

 with the following: 

Table 6-8. High Test Current Ground Bond Source Limits 

Nominal 

Value 

Required Standard 

Calibrator/Multimeter 

Current/Voltage 

Uncertainty  

DC Test 

Current 

R

gbr

 Lower 

Limit 

[1]

  

Upper Limit 

[1]

 

25 m

 

 

  0.5% 

20 A 

 

R

disp

 - 5 m

 

R

disp

 + 5 m

 

50 m

 

 

  0.2 % 

10 A 

 

R

disp

 - 5 m

 

R

disp

 + 5 m

 

100 m

 

 

  0.1 % 

10 A 

 

R

disp

 - 5 m

 

R

disp

 + 5 m

 

330 m

 

 

  0.1 % 

5 A 

 

R

disp

 - 7 m

 

R

disp

 + 7 m

 

500 m

 

 

  0.1 % 

3 A 

 

R

disp

 - 8 m

 

R

disp 

+ 8 m

 

 

 

  0.1 % 

2 A 

 

R

disp 

- 10 m

 

R

disp

 + 10 m

 

1.8 

 

 

  0.1 % 

2 A 

 

R

disp 

- 18 m

 

R

disp

 + 18 m

 

[1] R

disp

 = Displayed Value 

Change #6 

On page 7-17, replace 

Figure 7-14

 with the following: 

280V
RMS
MAX

280V
RMS
MAX

CAT I

1000V
CAT II

600V

RMS MAX

20V PK

30A

RMS
MAX

20V PK

LO 

LO    - SENSE 

, HI    , mA     

V

1500V PK

MAX

50V PK

MAX

20V PK

20V PK

N

L

PE

L1

L2

L3

H

E

S

OUTPUT

 ZL, Z GND, RCD

HI

LO

HI

LO

METER

METER    

5320A

 MULTIFUNCTION ELECTRICAL TESTER CALIBR

V

A

COM

INPUT

6500 

APPLIANCE TESTER

Fluke 5320A

Fluke 6500

Line Pin

 

Ehq039.eps 

Figure 7-14. Touch Leakage Current Calibration on Fluke 6500 

Change #7 

On page 6-14, replace Step 4 with: 

4.

 

Set the multifunction calibrator to the same voltage as the voltage on the mains input of the 
Calibrator (5320A). 115 V or 230 V for example. Set the multifunction calibrator frequency to 
55 Hz.

 

Содержание 5320A

Страница 1: ...nt Issue 9 Print Date September 2006 Issue Date 2 16 Revision Date 2 10 08 Page Count 7 This supplement contains information necessary to ensure the accuracy of the above manual This manual is distributed as an electronic manual on the following CD ROM CD Title 5320A CD Rev Date 2 1 09 CD PN 2634346 ...

Страница 2: ...es an input resistance of 0 The Calibrator can accommodate testers that have either 0 input impedance or a finite input impedance The Calibrator has a setup parameter R multiplier input to accommodate different input impedances for insulation testers The R multiplier input has a default setting of 0 for the most common type of testers with a virtual ground sensing circuit For other testers with an...

Страница 3: ...Manual Supplement 5320A Users 2 11 09 Change 4 On page 6 11 replace Figure 6 4 with the following ewt090 eps Figure 6 4 Resistance Multiplier Verification Connectioons ...

Страница 4: ...m 1 Rdisp Displayed Value Change 6 On page 7 17 replace Figure 7 14 with the following 280V RMS MAX 280V RMS MAX CAT I 1000V CAT II 600V RMS MAX 20V PK 30A RMS MAX 20V PK LO LO SENSE HI mA V 1500V PK MAX 50V PK MAX 20V PK 20V PK N L PE L1 L2 L3 H E S OUTPUT ZL Z GND RCD HI LO HI LO METER METER 5320A MULTIFUNCTION ELECTRICAL TESTER CALIBR V A COM INPUT 65 00 AP PL IANC E TE ST ER Fluke 5320A Fluke ...

Страница 5: ...ays If the 5320A has been powered off for more than 30 days the relay cleaning procedure should be performed before usage The 5320A with serial numbers higher than xx548xxxx is equipped with an internal power source that provides current during the cleaning procedure For 5320As with serial numbers lower than 548xxxxxx an external power supply must be connected to the PE and N terminals See the nex...

Страница 6: ... least 20 V with 3 A of current To clean the relays 1 Remove all external connections to the 5320A front panel 2 Navigate to the Relay cleaning procedure as outlined above Do not start the procedure 3 Set the external power supply to 20 V and connect as shown to the PE and N terminals of the 5320A It is best to use an external analog linear supply source If a switching dc supply source is used two...

Страница 7: ...requirements for industrial electromagnetic wave equipment and the seller or user should take notice of it This equipment is intended for use in business environments and not to be used in homes USA FCC 47 CFR 15 subpart B This product is considered an exempt device per clause 15 103 Change 10 255 468 On page 1 9 under W Fuse Protection replace the existing content with RCD input 3 15 A 250 V Fast...

Страница 8: ...ace Table 6 2 with Table 6 2 Measurement Input Fuses Input Fuse Fluke Part No RCD F3 15H250 V 5 mm x 20 mm 2743508 Leakage Current F100mL150 V 5 mm x 20 mm 2743513 Meter T20H500 V 6 3 mm x 32 mm 4778086 Loop Line Impedance T4H500 V 6 3 mm x 32 mm 2743524 ...

Отзывы: