Manual Supplement 

 

5320A Users 

 

6 2/16 

Note 

For units with serial number lower than xx548xxxx, do not run the cleaning procedure 

without connecting the external power supply. When the cleaning procedure is performed 

without connecting the external power supply, the relay contact resistance may become 

even worse. Changing the polarity of the dc supply source is recommended from cleaning 

to cleaning to avoid one-way transport of contact material from one side of the contacts 

to the other.

 

Change #9, 

449 

On page 1-7, add the following to the 

Symbols

 table: 

Ã

 

Conforms to relevant South Korean EMC 
Standards. 

On page 1-9, add the following to the 

General Specifications

Electromagnetic Compatibility (EMC)

 

International ...................................... IEC 61326-1: Basic Electromagnetic Environment 

CISPR 11: Group 1, Class A 

Group 1: Equipment has intentionally generated and/or uses conductively-coupled 
radio frequency energy that is necessary for the internal function of the equipment 
itself. 

Class A: Equipment is suitable for use in all establishments other than domestic and 
those directly connected to a low-voltage power supply network that supplies 
buildings used for domestic purposes. There may be potential difficulties in ensuring 
electromagnetic compatibility in other environments due to conducted and radiated 
disturbances.  

Emissions that exceed the levels required by CISPR 11 can occur when the 
equipment is connected to a test object. 

Korea (KCC) ..................................... Class A Equipment (Industrial Broadcasting & Communication Equipment) 

Class A: Equipment meets requirements for industrial electromagnetic wave 
equipment and the seller or user should take notice of it. This equipment is intended 
for use in business environments and not to be used in homes. 

USA (FCC) ....................................... 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103. 

Change #10, 

255, 468 

On page 1-9, under 

W

 

Fuse Protection,

 replace the existing content with: 

RCD input ......................................... 3.15 A, 250 V, Fast (F3.15H250V – 5 mm x 20 mm) 

Meter amps (A) input ........................ 20 A, 500 V, Fast (T20H500 V – 6.3 mm x 32 mm) 

Loop/Line impedance input............... 4 A, 500 V, Time delay (T4H500 V – 6.3 mm x 32 mm) 

On page 4-6, following Table 4-2, replace the sentence with: 

An optional resistance multiplier is offered to increase the range of high resistances from 350 M

 to 

10 T

 and applicable dc test voltage to 5 KV. Resistance multiplier is based on a passive T-type 

resistance network. Nominal value of multiplication is 1000. Exact value depends on resistance 
multiplier calibration data. Input resistance of the multiplier is approximately-300 M

. Resistance 

multiplier can be applied for calibration of such UUTs which use virtual ground (current to voltage 
converter) as current sense terminal with input resistance. For UUTs with finite input resistance and 
non-virtual ground on sense terminal, the input resistance must be taken into consideration. For this 
purpose, the Calibrator is equipped with a parameter “R multiplier input”. Input resistance of sense 
terminal of UUT should be written here to perform calibration correctly. 

Note      

Input resistance of the sense terminal is different for different models of megohmmeter. 

Do not to set the parameter to the correct value. For megohmmeters with virtual ground, 

 must be written and saved here before calibration. 

Содержание 5320A

Страница 1: ...nt Issue 9 Print Date September 2006 Issue Date 2 16 Revision Date 2 10 08 Page Count 7 This supplement contains information necessary to ensure the accuracy of the above manual This manual is distributed as an electronic manual on the following CD ROM CD Title 5320A CD Rev Date 2 1 09 CD PN 2634346 ...

Страница 2: ...es an input resistance of 0 The Calibrator can accommodate testers that have either 0 input impedance or a finite input impedance The Calibrator has a setup parameter R multiplier input to accommodate different input impedances for insulation testers The R multiplier input has a default setting of 0 for the most common type of testers with a virtual ground sensing circuit For other testers with an...

Страница 3: ...Manual Supplement 5320A Users 2 11 09 Change 4 On page 6 11 replace Figure 6 4 with the following ewt090 eps Figure 6 4 Resistance Multiplier Verification Connectioons ...

Страница 4: ...m 1 Rdisp Displayed Value Change 6 On page 7 17 replace Figure 7 14 with the following 280V RMS MAX 280V RMS MAX CAT I 1000V CAT II 600V RMS MAX 20V PK 30A RMS MAX 20V PK LO LO SENSE HI mA V 1500V PK MAX 50V PK MAX 20V PK 20V PK N L PE L1 L2 L3 H E S OUTPUT ZL Z GND RCD HI LO HI LO METER METER 5320A MULTIFUNCTION ELECTRICAL TESTER CALIBR V A COM INPUT 65 00 AP PL IANC E TE ST ER Fluke 5320A Fluke ...

Страница 5: ...ays If the 5320A has been powered off for more than 30 days the relay cleaning procedure should be performed before usage The 5320A with serial numbers higher than xx548xxxx is equipped with an internal power source that provides current during the cleaning procedure For 5320As with serial numbers lower than 548xxxxxx an external power supply must be connected to the PE and N terminals See the nex...

Страница 6: ... least 20 V with 3 A of current To clean the relays 1 Remove all external connections to the 5320A front panel 2 Navigate to the Relay cleaning procedure as outlined above Do not start the procedure 3 Set the external power supply to 20 V and connect as shown to the PE and N terminals of the 5320A It is best to use an external analog linear supply source If a switching dc supply source is used two...

Страница 7: ...requirements for industrial electromagnetic wave equipment and the seller or user should take notice of it This equipment is intended for use in business environments and not to be used in homes USA FCC 47 CFR 15 subpart B This product is considered an exempt device per clause 15 103 Change 10 255 468 On page 1 9 under W Fuse Protection replace the existing content with RCD input 3 15 A 250 V Fast...

Страница 8: ...ace Table 6 2 with Table 6 2 Measurement Input Fuses Input Fuse Fluke Part No RCD F3 15H250 V 5 mm x 20 mm 2743508 Leakage Current F100mL150 V 5 mm x 20 mm 2743513 Meter T20H500 V 6 3 mm x 32 mm 4778086 Loop Line Impedance T4H500 V 6 3 mm x 32 mm 2743524 ...

Отзывы: