RHapid-Cal Humidity Generator
Product Setup
9
Fill the Product with Distilled Water
Before the Product is connected to mains power (see
Mains Voltage
), fill the
Product with 50 ml to 60 ml of clean, distilled water (see Figure 1).
1. Remove the fill cap from the front panel (see
Front Panel
).
2. Use the fill syringe and insert the extension tube over the fill inlet on the
Product.
3. Remove the plunger from the fill syringe and carefully pour water into the
syringe. Water can be forced into the Product with the plunger, or bring the
syringe to a higher level than the fill inlet and allow the water to flow into the
Product. Watch the water level indicator to gauge the fill level as the reservoir
is filled. The Product takes approximately two syringes of water.
Monitor the water level indicator to ensure the water level is between the Max
and Min lines.
Caution
To avoid damage to the Product, do not overfill the Product. If
the Product is overfilled, water may flood the chamber. Use the
drain port on the back of the Product to drain excess water. If
water is found within the chamber, it must be wiped dry with a
clean paper towel.
ibw009.eps
Figure 1. Fill the Water Level