background image

279 FC

Calibration Manual

10

2.

Use a dry gas for the purge. 

The gas should have a dew point below -15 °C. Fluke recommends nitrogen or dried air. If the gas contains any 
water vapor or ice, moisture will form on the target. Use a relatively low flow rate of 2.4 to 4.8 liters/min 

 

(5  CFH to 10 CFH) for the purge. 

Ice Buildup on the Target 

If a small amount of ice forms:

1.

Place the cover back on the target and allow the purge gas to sublime the ice. 

2.

If more ice forms or a purge is not available, change the instrument set-point to a temperature equal to or greater 
than 50 °C with the target uncovered. 

3.

Allow all ice to melt and all water to dry before returning target to use. 

W

Caution

DO NOT wipe the target.

IR Temperature (IR Camera) Below Dew Point

Note

For IR temperature settings below dew point, do not leave the target uncovered for 
more than 5 s, since it can cause ice to form on the target. Temperature measurements 
above dew point do not require the use of cover. 

To make measurements below the dew point: 

1.

Put the target cover in place. 

2.

Adjust the set-point to the desired temperature and allow the instrument to stabilize at that temperature. 

3.

Set IR calibrator emissivity setting to 0.95 to match the emissivity of the Multimeter IR camera settings.

4.

Turn the control knob on the Multimeter to 

IR CAMERA

.

5.

Warm up the Multimeter for 10 minutes.

6.

Remove the target cover when the target stabilizes. 

7.

Place the Multimeter 20 inches (51 cm) from the black body as shown in Figure 3.

8.

Compare the indication on the Multimeter display with the display reading limits in Table 3.

9.

Replace the target cover. 

IR Temperature (IR Camera) Above Dew Point

To make measurements above dew point.

1.

Remove the target cover.

2.

Adjust to the set-point temperature and allow the instrument to stabilize at that temperature. 

3.

Place the Multimeter 20 in (51 cm) from the black body. See Figure 3.

4.

Compare the indication on the Multimeter display with the display reading limits in Table 3.

If the display indication falls outside of the range shown in Table 3, repair is necessary.

Содержание 279 FC

Страница 1: ...016 Fluke Corporation All rights reserved Specifications are subject to change without notice All product names are trademarks of their respective companies 279 FC True rms Thermal Multimeter Calibrat...

Страница 2: ...r nearest Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB...

Страница 3: ...6 MIN MAX Recording 6 Infrared Camera 6 Required Equipment 7 Performance Tests 7 Display Test 7 Backlight Test 7 Keypad Test 8 Function Performance Tests 8 Non Current and Non Temperature Test 8 Curr...

Страница 4: ...ii 279 FC Calibration Manual...

Страница 5: ...ow to use the Product Contact Fluke To contact Fluke call one of the following telephone numbers Technical Support USA 1 800 44 FLUKE 1 800 443 5853 Calibration Repair USA 1 888 99 FLUKE 1 888 993 585...

Страница 6: ...aged Examine the test leads for damaged insulation exposed metal or if the wear indicator shows Check test lead continuity The battery door must be closed and locked before you operate the Product Do...

Страница 7: ...tegory II is applicable to test and measuring circuits connected directly to utilization points socket outlets and similar points of the low voltage MAINS installation Measurement Category III is appl...

Страница 8: ...ttery IEC 62133 Electromagnetic Compatibility EMC International IEC 61326 1 Portable Electromagnetic Environment CISPR 11 Group 1 Class A IEC 61326 2 2 Group 1 Equipment has intentionally generated an...

Страница 9: ...200 Hz to 500 Hz 600 0 mV 0 1 mV 1 0 3 6 000 V 0 001 V 1 0 3 4 0 3 5 15 0 3 5 60 00 V 0 01 V 600 0 V 0 1 V 1000 V 1 V 1 All ac voltage ranges are specified from 1 of range to 100 of range 2 Crest fact...

Страница 10: ...tores up to 100 images Image Transfer Fluke Connect SmartView File Format is2 Range Resolution Measurement 1 99 99 Hz 0 01 Hz 0 1 1 999 9 Hz 0 1 Hz 0 1 1 1 Frequency is specified up to 500 Hz Minimum...

Страница 11: ...rify that the display operates correctly 1 Turn on the Multimeter 2 Verify that all the icons show correctly on the display If not repair is necessary See Contact Fluke on page 1 Backlight Test To ver...

Страница 12: ...arm up the Calibrator as necessary Refer to the Calibrator specifications for warm up time 4 Let the temperature of the Multimeter become stable to room temperature Non Current and Non Temperature Tes...

Страница 13: ...hanges the target emissivity and apparent temperature If any ice or moisture even though it is a slight amount is on the target the display temperature accuracy is invalid In addition ice buildup can...

Страница 14: ...int do not require the use of cover To make measurements below the dew point 1 Put the target cover in place 2 Adjust the set point to the desired temperature and allow the instrument to stabilize at...

Страница 15: ...True rms Thermal Multimeter Performance Tests 11 Figure 3 IR Camera Performance Test Configuration Fluke 4180 4181 Fluke 279 FC iFlex 20 in 51 cm...

Страница 16: ...50 V 49 94 50 06 10 V 9 97 10 03 500 V 500 6 499 4 1000 V 997 1003 DC Millivolts 3 mV 2 8 3 2 mV dc 500 mV 499 4 500 6 500 mV 500 6 499 4 AC Millivolts 30 mV 60 Hz 29 4 30 6 mV ac 500 mV 60 Hz 494 7 5...

Страница 17: ...d the cursor automatically moves to the next number 3 To enter 1234 push and 4 Push Start to go to the first calibration point C 01 If you enter the correct password C 01 shows on the display If the p...

Страница 18: ...d to expose the battery door 4 Turn the battery door latch until the unlock symbol aligns with the arrow 5 Lift off the battery door tilt stand 6 Remove the battery pack 7 Remove the holster starting...

Страница 19: ...o accept the value and proceed to the next step 5 After you push wait until the step number advances before you change the Calibrator source or turn the Mulitimeter control knob An adjustment step can...

Страница 20: ...s and battery packs near heat or fire Do not put in sunlight Clean the case with a damp cloth and weak detergent Do not use a solvent or cleaners with abrasives Dirt or moisture in the terminals can c...

Страница 21: ...ed 1 Place the battery pack into the battery compartment 2 Install the battery door tilt stand 3 Turn the battery door latch until the locked symbol aligns with the arrow The display shows an icon for...

Страница 22: ...00 mAh Rechargeable Lithium Ion Battery 4398817 Battery Sponge 4381848 Battery Door 4693466 1 2 turn Fastener 2278155 Case Bottom Screws 5 14 x 3 4 qty 4 832246 USB Connector Cover 4724737 Holster 469...

Отзывы: